FDA Adverse Event Malfunction Summary report: N

BD INTIMA-II 20GAX1.16IN PRN SLM NPVC

MDR report key: 13197237 · Received January 7, 2022

Report

Report Number
3014704491-2021-00422
Event Type
Malfunction
Date Received
January 7, 2022
Date of Event
October 12, 2021
Report Date
January 23, 2022
Manufacturer
BECTON DICKINSON MEDICAL DEVICES CO., LTD. SUZHOU
Product Code
FOZ
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: NO SAMPLE OR PHOTO WAS RECEIVED FOR EVALUATION. A DEVICE HISTORY REVIEW WAS CONDUCTED FOR LOT NUMBER 1111460 . OUR RECORDS SHOW THAT THIS IS THE ONLY INSTANCE OF THIS ISSUE OCCURRING IN THIS PRODUCTION BATCH. ACCORDING TO THE SAMPLING PLAN APPLIED FOR PRODUCT PERFORMANCE, THIS LOT WAS ACCEPTED AND RELEASED WITHOUT DEFECTS BEING NOTED DURING THE FINAL ASSEMBLY OR VISUAL INSPECTIONS. THE BD INTIMA-II IS AN INFUSION ONLY DEVICE AND IS NOT RATED FOR HIGH PRESSURE INJECTIONS. 2 RETAINED SAMPLES OF THIS LOT WERE TAKEN FOR HIGH-PRESSURE BLASTING TEST, ALL PASSED THE TEST, THERE WAS NO LEAKAGE BETWEEN THE CONNECTION SITE AND HIGH PRESSURE CONNECTOR H3 OTHER TEXT : SEE H.10.

Description of Event or Problem · 0

IT WAS REPORTED BD INTIMA-II 20GAX1.16IN PRN SLM NPVC HAD LEAKAGE ISSUES. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM (B)(6): "DURING CT ENHANCEMENT, FLUID LEAKAGE WAS FOUND AT THE TRANSFER OF INDWELLING NEEDLE."

Description of Event or Problem · 0

IT WAS REPORTED BD INTIMA-II 20GAX1.16IN PRN SLM NPVC HAD LEAKAGE ISSUES. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM CHINESE: "DURING CT ENHANCEMENT, FLUID LEAKAGE WAS FOUND AT THE TRANSFER OF INDWELLING NEEDLE...".

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2516992 BD INTIMA-II 20GAX1.16IN PRN SLM NPVC INTRAVASCULAR CATHETER FOZ BECTON DICKINSON MEDICAL DEVICES CO., LTD. SUZHOU 1111460

Patients

Seq Age Sex Outcome Treatment
1 Unknown