PUMP MMT-1712KL 640G V4.10 BK SF MG
Report
- Report Number
- 2032227-2022-101730
- Event Type
- Malfunction
- Date Received
- January 7, 2022
- Date of Event
- January 6, 2022
- Report Date
- July 13, 2022
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- UDI-DI
- 000000763000367046
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IS
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
(B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE 640G INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.
RETAINER RING = BLACK. ON 01/06/2022 THE CUSTOMER REPORTED THAT THE UNIT IS NOT WORKING. INSERTED A TEST P-CAP INTO THE RETAINER RING, AND IT LOCKED IN PLACE PROPERLY. THE FOLLOWING WERE NOTED DURING VISUAL INSPECTION: CRACKED BATTERY TUBE THREADS, CRACKED CASE ON THE BATTERY TUBE SIDE STARTING AT THE LEFT BELT CLIP RAIL, SCRATCHED CASE. DEVICE PASSED DISPLACEMENT, SLEEP CURRENT MEASUREMENT, ACTIVE CURRENT MEASUREMENT, AND SELF TESTS. NO UNEXPECTED BLANK DISPLAYS WERE NOTED DURING TESTING. THUS SOFTWARE WAS UTILIZED AND UPLOADED TRACE/HISTORY FILES PROPERLY. NO PUMP ERROR 25 WAS NOTED DURING TESTING, IN THE PUMP HISTORY FILE, IN THE LONG TRACE FILE, OR IN THE POWER MANAGEMENT GRAPH. THE ADAPT TOOL CONFIRMS THAT THE FOLLOWING BATTERY ALARMS/ALERTS OCCURRED AROUND THE EVENT DATE: 58=FAILED BATT TEST OCCURRED 01/01/2022 12:30:52. THE ADAPT TOOL ALSO CONFIRMS THAT PUMP ERROR 4 OCCURRED 01/04/2022 09:18:09. THE CASE WAS CUT OPEN. AFTER VISUAL INSPECTION, THERE IS CORROSION IN/AROUND THE FOLLOWING: PCBA2--J1, LCD FLEX CABLE TERMINAL. IN SUMMARY, THE CUSTOMER¿S REPORT THAT THE UNIT IS NOT WORKING WAS NOT CONFIRMED DURING TESTING. THE PUMP ERROR 4 IS DUE TO A HARDWARE PROBLEM. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE PUMP WAS NOT WORKING AS IT SHOULD BE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1777514 | PUMP MMT-1712KL 640G V4.10 BK SF MG | PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-1712KL | HG4UR1S | 000000763000367046 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |