ELEVATE¿ SPINAL SYSTEM
Report
- Report Number
- 1030489-2022-00028
- Event Type
- Malfunction
- Date Received
- January 7, 2022
- Date of Event
- December 17, 2021
- Report Date
- May 30, 2022
- Manufacturer
- WARSAW ORTHOPEDICS
- Product Code
- MAX
- UDI-DI
- 00643169430433
- PMA / PMN Number
- K172199
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
PRODUCT ANALYSIS OF PART#7770723 ; LOT# 0866922W ANALYSIS SUMMARY: VISUAL AND OPTICAL INSPECTION REVEALED THE PEEK PORTION HAS BROKEN AWAY FROM THE TITANIUM PORTION. THE SCREW THAT CONTROLS THE HEIGHT OF THE CAGE HAS BEEN THREADED IN TO FAR CAUSING THE CAGE TO BREAK. THIS TYPE OF DAMAGE IS CONSISTENT WITH EXCESSIVE FORCE APPLIED THE THREADING SCREW WHEN EXPANDING THE IMPLANT. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING AN EVENT HAPPENED DURING INTRA-OP OF THE REPORTED PRODUCTS. THE PRE-OP DIAGNOSIS WAS MENTIONED AS L5-S1 SPONDYLOLISTHESIS. IT WAS REPORTED THAT, TWO CAGES WERE BROKEN DURING THIS PROCEDURE. ONE OF THE CAGE WAS REMOVED FROM PATIENT BODY L2-3 LEVEL WITHOUT ANY FRAGMENT LEFT AND UNABLE TO REMOVE THE TITANIUM PORTION OF OTHER CAGE. TITANIUM POTION WAS LEFT IN THE PATIENT AT THE L5-S1 DISC SPACE. NO REVISION PLANNED AT THIS TIME. THERE WAS A DELAY OF 15-20 MINUTES REPORTED AS A RESULT OF PRODUCT MALFUNCTION. NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2476230 | ELEVATE¿ SPINAL SYSTEM | INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, | MAX | WARSAW ORTHOPEDICS | 7770723 | 0866922W | 00643169430433 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 66 YR | Male |