FDA Adverse Event Other Summary report: N

PHOENIX

MDR report key: 1317918 · Received December 23, 2008

Report

Report Number
2087532-2008-00100
Event Type
Other
Date Received
December 23, 2008
Date of Event
November 10, 2008
Report Date
November 10, 2008
Manufacturer
GAMBRO DASCO
Product Code
FII
Product Problem
Yes
Report Source
Distributor report
Reporter Location
NY, US
Reporter Occupation
INVALID DATA

Narratives

Description of Event or Problem · 1

GAMBRO RECEIVED A REPORT THAT DURING A HEAT CITRIC DISINFECTION PROCEDURE, THE BICARBONATE CONCENTRATE TUBING CAME OFF THE BLUE MALE CONCENTRATE CONNECTOR ASSEMBLY ON THE LOWER FRONT PANEL OF A PHOENIX MACHINE. FOLLOWING THIS DISCONNECTION, A NURSE WAS BURNED BY HOT WATER THAT SPRAYED ONTO HER LEG. THE NURSE WAS TREATED ACCORDING TO THE UNIT'S POLICY (SHE WAS NOT TREATED IN THE EMERGENCY ROOM OR SEEN BY A PHYSICIAN). THE NURSE RETURNED TO WORK THE FOLLOWING DAY. THE PHOENIX MACHINE REMAINS IN SERVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PHOENIX HEMODIALYSIS EQUIPMENT FII GAMBRO DASCO PHOENIX NA

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN Other