FDA Adverse Event Malfunction Summary report: N

BD¿ ALCOHOL SWABS

MDR report key: 13179081 · Received January 6, 2022

Report

Report Number
1920898-2021-01375
Event Type
Malfunction
Date Received
January 6, 2022
Date of Event
December 8, 2021
Report Date
February 16, 2022
Manufacturer
BD MEDICAL - DIABETES CARE
Product Code
LKB
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

DATE OF EVENT: UNKNOWN. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

H6: INVESTIGATION SUMMARY CUSTOMER RETURNED 2 OPENED AND 20 UNOPENED ALCOHOL SWAB PACKETS FROM LOT 1221802. THE 2 OPENED ALCOHOL SWAB PACKETS FEATURE ALCOHOL SWABS WITH SIGNIFICANT YELLOWISH DISCOLORATION. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH #1221802. ALL INSPECTIONS WERE PERFORMED PER THE APPLICABLE OPERATIONS QC SPECIFICATIONS. THERE WERE ZERO (0) NOTIFICATIONS NOTED THAT PERTAINED TO THE COMPLAINT. BASED ON THE SAMPLES RECEIVED, BD WAS ABLE TO CONFIRM THE CUSTOMER¿S INDICATED FAILURE OF SWAB DISCOLORATION. ROOT-CAUSE CANNOT BE DETERMINED. H3 OTHER TEXT : SEE H10.

Description of Event or Problem · 0

IT WAS REPORTED BD¿ ALCOHOL SWABS HAD FOREIGN MATERIAL ON THEM. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "CONSUMER REPORTED THAT THE ALCOHOL SWABS ARE DISCOLORED, SHOWING A YELLOWISH COLOR."

Description of Event or Problem · 0

IT WAS REPORTED BD¿ ALCOHOL SWABS HAD FOREIGN MATERIAL ON THEM. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "CONSUMER REPORTED THAT THE ALCOHOL SWABS ARE DISCOLORED, SHOWING A YELLOWISH COLOR."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
30998 BD¿ ALCOHOL SWABS PAD, ALCOHOL, DEVICE DISINFECTANT LKB BD MEDICAL - DIABETES CARE 1221802

Patients

Seq Age Sex Outcome Treatment
1 Unknown