Narratives

Additional Manufacturer Narrative · 0

(B)(4). SOURCE: (B)(6). CONCOMITANT MEDICAL PRODUCTS: CATALOG NUMBER: 010000913 LOT NUMBER:6903204 BRAND NAME: G7 HI-WALL E1 LINER. CATALOG NUMBER:010000659 LOT NUMBER: 6950183 BRAND NAME: G7 PPS LTD ACET SHELL. MULTIPLE REPORTS WERE SUBMITTED ALONG WITH THIS REPORT: 0001825034-2022-00036, 0001825034-2022-00037. CUSTOMER HAS INDICATED THAT THE PRODUCT IS IN PROCESS OF BEING RETURNED TO ZIMMER BIOMET FOR INVESTIGATION. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED.

Additional Manufacturer Narrative · 0

THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. VISUAL EXAMINATION OF THE RETURNED SHELL FOUND LIGHT SCRATCHES ON THE INNER RADIUS AND OUTER RIM OF THE SHELL. FOREIGN DEBRIS IS VISIBLE ON THE OUTER SHELL. NO DAMAGE WAS OBSERVED TO THE LOCKING GROOVE OR ORIENTATION FEATURES. DHR WAS REVIEWED AND NO DISCREPANCIES WERE FOUND. THE ROOT CAUSE IS UNABLE TO BE DETERMINED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING AN INITIAL HIP ARTHROPLASTY, THE LINER WOULD NOT SEAT INTO THE SHELL CAUSING A 35 MINUTES DELAY. NO ADVERSE EVENTS HAVE BEEN REPORTED AS A RESULT OF THE MALFUNCTION. ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE.

Description of Event or Problem · 0

NO ADDITIONAL EVENT INFORMATION TO REPORT AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
666270 G7 PPS LTD ACET SHELL 44A PROSTHESIS, HIP JDI ZIMMER BIOMET, INC. N/A 6950183 00880304524187

Patients

Seq Age Sex Outcome Treatment
1 76 YR Female SEE H10 NARRATIVE