FDA Adverse Event Malfunction Summary report: N

FUSION CYTOLOGY BRUSH

MDR report key: 13175071 · Received January 6, 2022

Report

Report Number
1037905-2022-00009
Event Type
Malfunction
Date Received
January 6, 2022
Date of Event
December 9, 2021
Report Date
January 6, 2022
Manufacturer
COOK ENDOSCOPY
Product Code
FDX
UDI-DI
10827002315256
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

510K- K192908. INVESTIGATION EVALUATION: A PRODUCT EVALUATION WAS NOT PERFORMED IN RESPONSE TO THIS REPORT BECAUSE THE PRODUCT SAID TO BE INVOLVED WAS NOT PROVIDED TO COOK FOR EVALUATION. THE REPORT COULD NOT BE CONFIRMED. THE DEVICE HISTORY RECORD FOR THE LOT NUMBER SAID TO BE INVOLVED WAS REVIEWED. A DISCREPANCY OR ANOMALY WAS NOT OBSERVED WITH THE PRODUCT THAT WAS RELEASED FOR DISTRIBUTION. INVESTIGATION CONCLUSION: WE COULD NOT CONDUCT A COMPLETE INVESTIGATION BECAUSE THE PRODUCT SAID TO BE INVOLVED WAS NOT RETURNED FOR EVALUATION. A DEFINITIVE CAUSE FOR THE REPORTED OBSERVATION COULD NOT BE DETERMINED. THE INSTRUCTIONS FOR USE STATES, "INSPECT THE DEVICE WITH PARTICULAR ATTENTION TO KINKS, BENDS, OR BREAKS. IF AN ABNORMALITY IS DETECTED THAT WOULD PROHIBIT PROPER WORKING CONDITIONS, DO NOT USE." PRIOR TO DISTRIBUTION, ALL FUSION CYTOLOGY BRUSHES ARE SUBJECTED TO A VISUAL INSPECTION AND FUNCTIONAL TESTING TO ENSURE DEVICE INTEGRITY. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THAT THE LOTS SAID TO BE INVOLVED MET ALL MANUFACTURING REQUIREMENTS PRIOR TO SHIPMENT. CORRECTIVE ACTION: A REVIEW OF THE COMPLAINT HISTORY WAS CONDUCTED AND THIS REPRESENTS AN ISOLATED OCCURRENCE. THE LIKELIHOOD OF OCCURRENCE IS CONSIDERED REMOTE. CORRECTIVE ACTION IS NOT WARRANTED AT THIS TIME BASED ON THE QUALITY ENGINEERING RISK ASSESSMENT. QUALITY ASSURANCE WILL CONTINUE TO MONITOR FOR COMPLAINT TRENDS AND REASSESS THE RISK ASSESSMENT RESULTS AS POST MARKET FEEDBACK CONTINUES TO BECOME AVAILABLE.

Description of Event or Problem · 0

IN PREPARATION FOR A CYTOLOGY PROCEDURE, THE USER SELECTED A COOK FUSION CYTOLOGY BRUSH,. THE BRUSH WOULD NOT RETRACT INTO THE CATHETER AND FROM TRYING TO RETRACT, THE BRUSH BROKE OFF, SUBJECT OF REPORT. THIS OCCURRED PRIOR TO PATIENT CONTACT; THERE WAS NO IMPACT TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
74289 FUSION CYTOLOGY BRUSH FDX ENDOSCOPIC CYTOLOGY BRUSH FDX COOK ENDOSCOPY W4513342 10827002315256

Patients

Seq Age Sex Outcome Treatment
1 Unknown