FDA Adverse Event Malfunction Summary report: N

BD VERITOR ¿ SARS-COV-2 & FLU A+B

MDR report key: 13170441 · Received January 5, 2022

Report

Report Number
3006948883-2021-01045
Event Type
Malfunction
Date Received
January 5, 2022
Date of Event
December 9, 2021
Report Date
March 9, 2022
Manufacturer
BD RAPIS DIAGNOSTICS (SUZHOU) CO. LTD.
Product Code
QMN
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D10: DEVICE AVAILABLE FOR EVAL: YES. D10: RETURNED TO MANUFACTURER ON: 2021-12-22. H6: INVESTIGATION SUMMARY: THIS STATEMENT IS TO SUMMARIZE THE INVESTIGATION RESULTS REGARDING THE COMPLAINT THAT ALLEGES FALSE NEGATIVE WHEN USING KIT RAPID DETECTION OF SARS-COV-2 VERITOR (MATERIAL # 256082), BATCH NUMBER 1193294. BD QUALITY PERFORMS A SYSTEMATIC APPROACH TO INVESTIGATE FALSE NEGATIVE COMPLAINTS. THIS APPROACH INVOLVES REVIEW OF MANUFACTURING BATCH HISTORY RECORDS, TESTING OF RETENTION SAMPLES, AND TESTING OF CUSTOMER RETURNED SAMPLES, IF APPLICABLE. AN INVESTIGATION AND TESTING WERE PERFORMED ON THE BATCH NUMBER PROVIDED. RETURNED PRODUCT TESTING WAS COMPLETED AND ALL DEVICES HAD NORMAL INTENDED RESULTS. THE REPORTED ISSUE WAS UNABLE TO BE CONFIRMED. THE ROOT CAUSE COULD NOT BE IDENTIFIED. BD QUALITY WILL CONTINUE TO CLOSELY MONITOR FOR TRENDS.

Additional Manufacturer Narrative · 0

EUA# (B)(4). A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE TESTING WITH BD VERITOR ¿ SARS-COV-2 & FLU A+B, 5 FALSE NEGATIVE RESULTS WERE OBTAINED. CONFIRMATORY PCR TESTING WAS PERFORMED, AND THE RESULTS WERE POSITIVE. THERE WAS NO PATIENT IMPACT. EUA# (B)(4). THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER REPORTS FALSE NEGATIVE RESULTS.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE TESTING WITH BD VERITOR ¿ SARS-COV-2 & FLU A+B, 5 FALSE NEGATIVE RESULTS WERE OBTAINED. CONFIRMATORY PCR TESTING WAS PERFORMED, AND THE RESULTS WERE POSITIVE. THERE WAS NO PATIENT IMPACT. EUA# (B)(4). THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER REPORTS FALSE NEGATIVE RESULTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1133296 BD VERITOR ¿ SARS-COV-2 & FLU A+B NOT CLASSIFIED QMN BD RAPIS DIAGNOSTICS (SUZHOU) CO. LTD. 1193294

Patients

Seq Age Sex Outcome Treatment
1 Unknown