FDA Adverse Event Malfunction Summary report: N

HEATER-COOLER SYSTEM 3T

MDR report key: 13169468 · Received January 5, 2022

Report

Report Number
9611109-2022-00008
Event Type
Malfunction
Date Received
January 5, 2022
Date of Event
September 29, 2021
Report Date
March 4, 2022
Manufacturer
LIVANOVA DEUTSCHLAND GMBH
Product Code
DWC
PMA / PMN Number
K191402
Removal / Correction Number
Z-2076/2081-2015
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

THERE WAS NO KNOWN PATIENT INVOLVEMENT. THE HEATER-COOLER 16-02-80 IS NOT DISTRIBUTED IN THE USA AND IT IS SIMILAR TO HEATER-COOLER 16-02-85, WHICH IS DISTRIBUTED IN THE USA (510(K) NUMBER: K191402). LIVANOVA (B)(4) IMPLEMENTED A FIELD SAFETY NOTICE FOR DISINFECTION AND CLEANING OF HEATER-COOLER DEVICES. THE Z NUMBER IS Z-2076/2081-2015. LIVANOVA (B)(4) MANUFACTURES THE HEATER-COOLER SYSTEM 3T . THE INCIDENT OCCURRED IN (B)(6). LIVANOVA INITIATED AN INVESTIGATION. IF ANY ADDITIONAL INFORMATION PERTINENT TO THE REPORTED EVENT IS RECEIVED, IT WILL BE PROVIDED IN A SUPPLEMENTAL REPORT.

Additional Manufacturer Narrative · 0

THERE WAS NO KNOWN PATIENT INVOLVEMENT. THE HEATER-COOLER 16-02-80 IS NOT DISTRIBUTED IN THE USA AND IT IS SIMILAR TO HEATER-COOLER 16-02-85, WHICH IS DISTRIBUTED IN THE USA (510(K) NUMBER: K191402). LIVANOVA (B)(4) IMPLEMENTED A FIELD SAFETY NOTICE FOR DISINFECTION AND CLEANING OF HEATER-COOLER DEVICES. THE Z NUMBER IS Z-2076/2081-2015. LIVANOVA (B)(4) MANUFACTURES THE HEATER-COOLER SYSTEM 3T . THE INCIDENT OCCURRED IN (B)(6). LIVANOVA INITIATED AN INVESTIGATION. IF ANY ADDITIONAL INFORMATION PERTINENT TO THE REPORTED EVENT IS RECEIVED, IT WILL BE PROVIDED IN A SUPPLEMENTAL REPORT.

Additional Manufacturer Narrative · 0

THROUGH PREVIOUS FOLLOW-UP COMMUNICATION FROM THE SAME HOSPITAL AND FOR THE SAME UNIT, LIVANOVA LEARNED THAT THE DEVICE WAS CLEANED REGULARLY PER THE INSTRUCTION FOR USE AND THAT IT WAS PLACED INSIDE THE OPERATING THEATER DURING USE WITH THE FAN DIRECTED AWAY FROM THE PATIENT AND AT AN ESTIMATED DISTANCE OF 3-4 METERS FROM THE SURGERY FIELD. NO ADDITIONAL INFORMATION IS AVAILABLE ABOUT THIS SPECIFIC CASE AND THE ROOT CAUSE COULD NOT BE DETERMINED.

Description of Event or Problem · 0

LIVANOVA RECEIVED A REPORT OF HEATER COOLER 3T CONTAMINATED WITH NTM CHIMAERA.NO PATIENT INVOLVEMENT REPORTED.

Description of Event or Problem · 0

LIVANOVA RECEIVED A REPORT OF HEATER COOLER 3T CONTAMINATED WITH NTM CHIMAERA.NO PATIENT INVOLVEMENT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1436338 HEATER-COOLER SYSTEM 3T CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS DWC LIVANOVA DEUTSCHLAND GMBH 16-02-80

Patients

Seq Age Sex Outcome Treatment
1 Unknown