FDA Adverse Event Malfunction Summary report: N

BD¿ SLIP-TIP SYRINGE WITH ATTACHED NEEDLE

MDR report key: 13160045 · Received January 4, 2022

Report

Report Number
1213809-2021-00835
Event Type
Malfunction
Date Received
January 4, 2022
Date of Event
December 3, 2021
Report Date
January 6, 2022
Manufacturer
BECTON DICKINSON MEDICAL SYSTEMS
Product Code
FMF
UDI-DI
30382903095972
PMA / PMN Number
K110771
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

H6: INVESTIGATION SUMMARY . TWO PHOTOS WERE RECEIVED AND EVALUATED. ONE PHOTO DISPLAYED THE GRAPHIC SIDE OF A BLISTER PACKAGE (P/N 309597) FROM BATCH #0150719. THE SECOND PHOTO DISPLAYED A 1ML SYRINGE WITH NEEDLE IN ITS BLISTER PACKAGE WITH NEEDLE ATTACHED. AN ARROW POINTING TO THE NEEDLE/TIP CONNECTION WAS PRESENT. THE CONNECTION DID NOT APPEAR TO BE IRREGULAR FROM THE PHOTO AND NO NON-CONFORMITIES WERE OBSERVED. UNUSED PHYSICAL SAMPLES ARE NEEDED TO PERFORM A MORE IN-DEPTH EVALUATION. SINCE NO SAMPLES DISPLAYING THE REPORTED CONDITION WERE RECEIVED A POTENTIAL ROOT CAUSE COULD NOT BE DEFINED AND CORRECTIVE ACTIONS ARE NOT NECESSARY. A DEVICE HISTORY RECORD REVIEW SHOWED NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION OF THE PROVIDED LOT NUMBER THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT. H3 OTHER TEXT : SEE H10.

Description of Event or Problem · 0

IT WAS REPORTED THAT 5 BD¿ SLIP-TIP SYRINGES WITH ATTACHED NEEDLE PLUNGERS WERE DIFFICULT TO PUSH DURING USE. THIS COMPLAINT WAS CREATED TO CAPTURE THE 2ND OF 2 RELATED INCIDENTS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "SPOKE WITH PATIENT'S DAUGHTER COMPLAINTS THAT PLUNGER ON INJECTION SYRINGES ARE SOMETIMES HARD TO PUSH, STATES IT OCCURS IN ABOUT FIVE SYRINGES OUT OF A MONTH SUPPLY."

Description of Event or Problem · 0

IT WAS REPORTED THAT 5 BD¿ SLIP-TIP SYRINGES WITH ATTACHED NEEDLE PLUNGERS WERE DIFFICULT TO PUSH DURING USE. THIS COMPLAINT WAS CREATED TO CAPTURE THE 2ND OF 2 RELATED INCIDENTS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "SPOKE WITH PATIENT'S DAUGHTER COMPLAINTS THAT PLUNGER ON INJECTION SYRINGES ARE SOMETIMES HARD TO PUSH, STATES IT OCCURS IN ABOUT FIVE SYRINGES OUT OF A MONTH SUPPLY."

Description of Event or Problem · 0

IT WAS REPORTED THAT 5 BD¿ SLIP-TIP SYRINGES WITH ATTACHED NEEDLE PLUNGERS WERE DIFFICULT TO PUSH DURING USE. THIS COMPLAINT WAS CREATED TO CAPTURE THE 2ND OF 2 RELATED INCIDENTS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "SPOKE WITH PATIENT'S DAUGHTER COMPLAINTS THAT PLUNGER ON INJECTION SYRINGES ARE SOMETIMES HARD TO PUSH, STATES IT OCCURS IN ABOUT FIVE SYRINGES OUT OF A MONTH SUPPLY."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
479506 BD¿ SLIP-TIP SYRINGE WITH ATTACHED NEEDLE PISTON SYRINGE FMF BECTON DICKINSON MEDICAL SYSTEMS 309597 0150719 30382903095972

Patients

Seq Age Sex Outcome Treatment
1 Unknown