FDA Adverse Event Malfunction Summary report: N

BD LUER-LOK¿ TIP SYRINGE

MDR report key: 13156351 · Received January 4, 2022

Report

Report Number
1213809-2021-00831
Event Type
Malfunction
Date Received
January 4, 2022
Date of Event
December 6, 2021
Report Date
January 14, 2022
Manufacturer
BECTON DICKINSON MEDICAL SYSTEMS
Product Code
FMF
UDI-DI
30382903096283
PMA / PMN Number
K941562
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

H6: INVESTIGATION SUMMARY ONE LOOSE 1ML SYRINGE (P/N 309628) WAS RECEIVED AND EVALUATED. THE SYRINGE HAD A PLUNGER ROD WITH A STOPPER AND A BD TIP CAP SEPARATED FROM IT. THE SYRINGE HAD A SKEWED SCALE. THE CONDITION OBSERVED WAS NON-CONFORMING PER PRODUCT SPECIFICATION. POTENTIAL ROOT CAUSE FOR THE SKEWED SCALE DEFECT IS ASSOCIATED WITH THE MARKING PROCESS. THESE CONDITIONS ARE OCCURRING AT/BELOW THEIR EXPECTED FREQUENCY. THEREFORE, NO CORRECTIVE ACTION IS REQUIRED AT THIS TIME. BATCH #1081816 IS CONSIDERED IN COMPLIANCE WITH OUR PRODUCT SPECIFICATION REQUIREMENTS. A DEVICE HISTORY RECORD REVIEW SHOWED NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION OF THE PROVIDED LOT NUMBER THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT. H3 OTHER TEXT : SEE H10.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD LUER-LOK¿ TIP SYRINGE HAD SCALE MARKING ISSUES. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "SCALE ERROR"

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD LUER-LOK¿ TIP SYRINGE HAD SCALE MARKING ISSUES. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "SCALE ERROR."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1160313 BD LUER-LOK¿ TIP SYRINGE PISTON SYRINGE FMF BECTON DICKINSON MEDICAL SYSTEMS 309628 1081816 30382903096283

Patients

Seq Age Sex Outcome Treatment
1 Unknown