BARD® EXTENSION CABLE
Report
- Report Number
- 1018233-2021-08628
- Event Type
- Malfunction
- Date Received
- December 31, 2021
- Date of Event
- December 16, 2021
- Report Date
- April 8, 2022
- Manufacturer
- C.R. BARD, INC. (COVINGTON) -1018233
- Product Code
- EZL
- UDI-DI
- 00801741045875
- PMA / PMN Number
- K070582
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
THE REPORTED EVENT WAS UNCONFIRMED BECAUSE THE REPORTED FAILURE COULD NOT BE REPRODUCED. A TEMPERATURE SENSING CATHETER EXTENSION CORD WAS RETURNED OPENED WITHOUT ORIGINAL PACKAGING. A PHOTO SAMPLE WAS ALSO RETURNED. NO DAMAGE TO THE CORD WAS NOTED. UNABLE TO DETERMINE IF THE CABLE HAD TEMPERATURE DISPLAY ISSUES FROM THE PHOTO SO THE PHYSICAL SAMPLE WAS EVALUATED. NO DAMAGE TO THE CORD WAS NOTED. CONNECTIONS OF THE CORD TO THE PLUG AND SOCKET WERE SECURE. USING A MULTIMETER, THE CORD WAS TESTED FOR AND FOUND TO HAVE CONTINUITY. THE CORD WAS CONNECTED TO AN IN-HOUSE TEMPERATURE SENSING CATHETER SUBMERGED IN A WATER BATH (119118 LOT NGEVZ327) THE CORD WAS THEN ATTACHED TO A KILO DEVICE AND WAS ABLE TO DISPLAY THE TEMPERATURE. TEMPERATURE DISPLAYED WAS SLIGHTLY HIGHER THAN THE THERMOMETER TEMPERATURE. PER PRODUCT QE, THIS DIFFERENCE WOULD NOT BE INDICATIVE OF A FAILURE. THE SAMPLE MEETS THE SPECIFICATION "PLUG AND JACK MUST BE FIRMLY SECURED TO WIRE". THE PRODUCT HAD NOT CAUSED THE REPORTED FAILURE. NO ROOT CAUSE COULD BE FOUND BECAUSE THE REPORTED EVENT WAS UNCONFIRMED. THE DEVICE WAS USED FOR DIAGNOSTIC PURPOSES. THE LOT NUMBER IS UNKNOWN; THEREFORE, THE DEVICE HISTORY RECORD COULD NOT BE REVIEWED. AS THE REPORTED EVENT IS UNCONFIRMED A LABELING REVIEW IS NOT REQUIRED. THE ACTUAL/SUSPECTED DEVICE WAS INSPECTED
THE INVESTIGATION IS STILL IN PROGRESS. ONCE THE INVESTIGATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT A SLIGHT TOUCH CAUSED AN ERRATIC DISPLAY OR A LOW TEMPERATURE DISPLAY. THERE WERE MANY OF THE SIMILAR PROBLEMS SINCE THE 153622-MODEL NUMBER WAS INTRODUCED. THE CUSTOMER COMPLAINED THAT THE STRENGTH OF THE CABLE HAD BEEN REDUCED.
IT WAS REPORTED THAT A SLIGHT TOUCH CAUSED AN ERRATIC DISPLAY OR A LOW TEMPERATURE DISPLAY. THERE WERE MANY OF THE SIMILAR PROBLEMS SINCE THE 153622-MODEL NUMBER WAS INTRODUCED. THE CUSTOMER COMPLAINED THAT THE STRENGTH OF THE CABLE HAD BEEN REDUCED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2027932 | BARD® EXTENSION CABLE | EXTENSION CABLE | EZL | C.R. BARD, INC. (COVINGTON) -1018233 | 153622 | BMFU0001 | 00801741045875 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | Other |