FDA Adverse Event Malfunction Summary report: N

BARD® EXTENSION CABLE

MDR report key: 13125594 · Received December 31, 2021

Report

Report Number
1018233-2021-08628
Event Type
Malfunction
Date Received
December 31, 2021
Date of Event
December 16, 2021
Report Date
April 8, 2022
Manufacturer
C.R. BARD, INC. (COVINGTON) -1018233
Product Code
EZL
UDI-DI
00801741045875
PMA / PMN Number
K070582
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE REPORTED EVENT WAS UNCONFIRMED BECAUSE THE REPORTED FAILURE COULD NOT BE REPRODUCED. A TEMPERATURE SENSING CATHETER EXTENSION CORD WAS RETURNED OPENED WITHOUT ORIGINAL PACKAGING. A PHOTO SAMPLE WAS ALSO RETURNED. NO DAMAGE TO THE CORD WAS NOTED. UNABLE TO DETERMINE IF THE CABLE HAD TEMPERATURE DISPLAY ISSUES FROM THE PHOTO SO THE PHYSICAL SAMPLE WAS EVALUATED. NO DAMAGE TO THE CORD WAS NOTED. CONNECTIONS OF THE CORD TO THE PLUG AND SOCKET WERE SECURE. USING A MULTIMETER, THE CORD WAS TESTED FOR AND FOUND TO HAVE CONTINUITY. THE CORD WAS CONNECTED TO AN IN-HOUSE TEMPERATURE SENSING CATHETER SUBMERGED IN A WATER BATH (119118 LOT NGEVZ327) THE CORD WAS THEN ATTACHED TO A KILO DEVICE AND WAS ABLE TO DISPLAY THE TEMPERATURE. TEMPERATURE DISPLAYED WAS SLIGHTLY HIGHER THAN THE THERMOMETER TEMPERATURE. PER PRODUCT QE, THIS DIFFERENCE WOULD NOT BE INDICATIVE OF A FAILURE. THE SAMPLE MEETS THE SPECIFICATION "PLUG AND JACK MUST BE FIRMLY SECURED TO WIRE". THE PRODUCT HAD NOT CAUSED THE REPORTED FAILURE. NO ROOT CAUSE COULD BE FOUND BECAUSE THE REPORTED EVENT WAS UNCONFIRMED. THE DEVICE WAS USED FOR DIAGNOSTIC PURPOSES. THE LOT NUMBER IS UNKNOWN; THEREFORE, THE DEVICE HISTORY RECORD COULD NOT BE REVIEWED. AS THE REPORTED EVENT IS UNCONFIRMED A LABELING REVIEW IS NOT REQUIRED. THE ACTUAL/SUSPECTED DEVICE WAS INSPECTED

Additional Manufacturer Narrative · 0

THE INVESTIGATION IS STILL IN PROGRESS. ONCE THE INVESTIGATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED THAT A SLIGHT TOUCH CAUSED AN ERRATIC DISPLAY OR A LOW TEMPERATURE DISPLAY. THERE WERE MANY OF THE SIMILAR PROBLEMS SINCE THE 153622-MODEL NUMBER WAS INTRODUCED. THE CUSTOMER COMPLAINED THAT THE STRENGTH OF THE CABLE HAD BEEN REDUCED.

Description of Event or Problem · 0

IT WAS REPORTED THAT A SLIGHT TOUCH CAUSED AN ERRATIC DISPLAY OR A LOW TEMPERATURE DISPLAY. THERE WERE MANY OF THE SIMILAR PROBLEMS SINCE THE 153622-MODEL NUMBER WAS INTRODUCED. THE CUSTOMER COMPLAINED THAT THE STRENGTH OF THE CABLE HAD BEEN REDUCED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2027932 BARD® EXTENSION CABLE EXTENSION CABLE EZL C.R. BARD, INC. (COVINGTON) -1018233 153622 BMFU0001 00801741045875

Patients

Seq Age Sex Outcome Treatment
1 Unknown Other