VERITAS VISION SYSTEM
Report
- Report Number
- 3012236936-2021-00385
- Event Type
- Malfunction
- Date Received
- December 30, 2021
- Date of Event
- December 8, 2021
- Report Date
- February 3, 2022
- Manufacturer
- JOHNSON & JOHNSON SURGICAL VISION, INC.
- Product Code
- HQC
- UDI-DI
- 05050474700864
- PMA / PMN Number
- K203060
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MS, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
H4 - MANUFACTURING DATE - THE MANUFACTURING SITE REPORTED THAT THE MANUFACTURING DATE FOR THE DEVICE IS 9/24/2021. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC. HAS BEEN SUBMITTED.
DEVICE EVALUATION: FIELD SERVICE AGENT TESTED THE SYSTEM AND WAS NOT ABLE TO DUPLICATE REPORTED ISSUE BUT DID SEE ERRORS IN EVENT LOG. VERITAS CHECKLIST AND SYSTEM CHECKS USING STANDARD SETTING WERE PERFORMED AND NO PROBLEM WAS FOUND. VERITAS MEETS SPECIFICATION. MANUFACTURING RECORD REVIEW: A REVIEW OF RECORDS RELATED TO THE DEVICE INCLUDING LABELING, COMPLAINT TRENDING, AND RISK DOCUMENTATION WILL BE PERFORMED. UPON COMPLETION OF THIS REVIEW, IF THERE IS ANY FURTHER RELEVANT INFORMATION OBTAINED, A SUPPLEMENTAL MEDWATCH WILL BE FILED. CONCLUSION: BASED ON THE INVESTIGATION RESULTS THERE IS NO INDICATION OF A PRODUCT QUALITY DEFICIENCY. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC. HAS BEEN SUBMITTED
IT WAS REPORTED THAT THE PHACO MACHINE GAVE160 ERROR DURING PRIME AND DURING QUADRANT. THE 160 ERROR REFERS TO ADVANCED INFUSION PRESSURE IS LOW. THE ERROR WAS OBSERVED TWICE ONE DURING PRIMING AND ANOTHER DURING THE PHACO PROCEDURE. THERE WAS NO PATIENT INJURY AND NO REPORTED SURGICAL INTERVENTION. NO FURTHER INFORMATION PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2020538 | VERITAS VISION SYSTEM | UNIT, PHACOFRAGMENTATION | HQC | JOHNSON & JOHNSON SURGICAL VISION, INC. | VRT680300 | 05050474700864 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |