FDA Adverse Event Malfunction Summary report: N

TRUE METRIX

MDR report key: 13119244 · Received December 30, 2021

Report

Report Number
1000113657-2021-00812
Event Type
Malfunction
Date Received
December 30, 2021
Date of Event
December 2, 2021
Report Date
December 30, 2021
Manufacturer
TRIVIDIA HEALTH INC
Product Code
NBW
UDI-DI
00021292007836
PMA / PMN Number
K140100
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

INTERNAL REPORT REFERENCE NUMBER: (B)(4). TEST STRIPS WERE NOT RETURNED FOR EVALUATION. METER WAS RETURNED - REPORTED DEFECT NOT REPRODUCED. NO DEFECT FOUND MOST LIKELY UNDERLYING ROOT CAUSE: (B)(4): USER HAD AN INACCURATE REFERENCE: SELF: THE PERSON IS USING THEMSELVES AS THE REFERENCE OR HOW THEY FEEL AT THE TIME THEY RUN THE BLOOD TEST. NOTE: MANUFACTURER CONTACTED CUSTOMER IN A FOLLOW-UP CALL TO ENSURE THE REPLACEMENT PRODUCTS RESOLVED THE INITIAL CONCERN - UNABLE TO ESTABLISH CONTACT WITH CUSTOMER AT THIS TIME.

Description of Event or Problem · 0

CONSUMER REPORTED COMPLAINT FOR HIGH AND ERRATIC BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TEST RESULTS FROM RESULTS OBTAINED OF 143, 191, 320, 313 AND 333MG/DL. THE CUSTOMER¿S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 120-140MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL, A BACK TO BACK BLOOD TEST WAS PERFORMED BY THE CUSTOMER NON-FASTING AND PRODUCED TEST RESULTS OF 291 MG/DL AND 242MG/DL USING TRUE METRIX METER; CUSTOMER WAS NOT SATISFIED WITH THE RESULTS OBTAINED. THE PRODUCT IS STORED ACCORDING TO SPECIFICATION IN THE BEDROOM. THE TEST STRIP LOT MANUFACTURER¿S EXPIRATION DATE IS 08/19/2022 AND OPEN VIAL DATE IS 11/2021. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY (DATE/TIME NOT SET): RESULT 1 : 143 MG/DL DATE: (B)(6) 2021 TIME: 5:00 PM FASTING, RESULT 2 : 191 MG/DL DATE: (B)(6) 2021 TIME: 4:59 PM FASTING, RESULT 3 : 320 MG/DL DATE: (B)(6) 2021 TIME: 8:15 PM FASTING, RESULT 4 : 313 MG/DL DATE: (B)(6) 2021 TIME: 8:13 AM FASTING, RESULT 5 : 333 MG/DL DATE: (B)(6) 2021 TIME: 8:13 AM FASTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2021110 TRUE METRIX SYSTEM, TEST BLOOD GLUCOSE, OVER THE COUNTER NBW TRIVIDIA HEALTH INC STRIP, SUNMARK TMX 50CTMG/DL MM ZX4267S 00021292007836

Patients

Seq Age Sex Outcome Treatment
1 Unknown