FDA Adverse Event Malfunction Summary report: N

ID NOW COVID-19 ASSAY

MDR report key: 13101753 · Received December 28, 2021

Report

Report Number
1221359-2021-03929
Event Type
Malfunction
Date Received
December 28, 2021
Report Date
January 30, 2022
Manufacturer
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
Product Code
QJR
UDI-DI
04571226475027
PMA / PMN Number
EUA2000074
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

THIS INVESTIGATION IS STILL IN PROGRESS. ONCE THE INVESTIGATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE PROVIDED. REFERENCE MFR. REPORT: 1221359-2021-03930.

Additional Manufacturer Narrative · 0

INVESTIGATION: TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT 1052056 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 190-000 / LOT 1052056 AND TEST BASE PART NUMBER 190-430 / LOT 1052056. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS CONFLICTING RESULTS PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 1052056 SHOWED THAT THE COMPLAINT RATE IS (B)(4). ABBOTT DIAGNOSTICS SCARBOROUGH WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE HOWEVER A POSSIBLE ASSIGNABLE ROOT CAUSE IS PATIENT SAMPLE INTERFERENCE.

Description of Event or Problem · 0

THE CUSTOMER REPORTED TWO (2) FCONFLICTING RESULTS WITH THE ID NOW COVID-19 ASSAY. THIS REPORT IS FOR PATIENT ONE (1) OF TWO (2). THE CUSTOMER REPORTED A CONFLICTING RESULT ON A DIRECTLY TESTED NASAL SWAB WITH THE ID NOW COVID-19 ASSAY. PER THE CUSTOMER, THERE WAS NO PATIENT HARM DUE TO TEST RESULTS. ADDITIONALLY, THERE WAS NO DELAY OR IMPACT IN TREATMENT DUE TO TEST RESULTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1997425 ID NOW COVID-19 ASSAY MOLECULAR IVD FOR ID NOW COVID-19 QJR ABBOTT DIAGNOSTICS SCARBOROUGH, INC. 1052056 04571226475027

Patients

Seq Age Sex Outcome Treatment
1 Unknown