FDA Adverse Event Malfunction Summary report: N

BD HYPODERMIC NEEDLE PRECISIONGLIDE¿

MDR report key: 13099348 · Received December 28, 2021

Report

Report Number
1911916-2021-01292
Event Type
Malfunction
Date Received
December 28, 2021
Date of Event
December 1, 2021
Report Date
December 11, 2021
Manufacturer
BD MEDICAL (BD WEST) MEDICAL SURGICAL
Product Code
FMI
UDI-DI
30382903051107
PMA / PMN Number
K021475
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED AS A DEFAULT. THE CUSTOMER'S ADDRESS IS UNKNOWN. (B)(6) USA HAS BEEN USED AS A DEFAULT. INVESTIGATION SUMMARY: A DEVICE HISTORY RECORD REVIEW WAS COMPLETED BY OUR QUALITY ENGINEER TEAM FOR PROVIDED MATERIAL NUMBER 305110 AND LOT NUMBER 0022962. THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THIS DEFECT AND ALL QUALITY TESTS WERE FOUND TO BE WITHIN SPECIFICATION. AS A SAMPLE WAS UNAVAILABLE FOR RETURN, A THOROUGH SAMPLE INVESTIGATION COULD NOT BE COMPLETED. BASED ON THE INVESTIGATION RESULTS, AN EXACT CAUSE FOR THIS INCIDENT COULD NOT BE IDENTIFIED. THERE ARE QUALITY CONTROLS CURRENTLY IN PLACE TO DETECT THIS TYPE OF DEFECT DURING THE PRODUCTION PROCESS. FURTHER ACTION HAS NOT BEEN DETERMINED NECESSARY AT THIS TIME. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR QUALITY TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD HYPODERMIC NEEDLE PRECISIONGLIDE¿ CONTAINED FOREIGN MATTER IN THE DEVICE CANNULA. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "IT WAS REPORTED BY THE CONSUMER WHITE MATERIAL WAS FOUND FLOATING IN THE INSULIN VIAL AFTER INSERTING THE NEEDLE."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1999311 BD HYPODERMIC NEEDLE PRECISIONGLIDE¿ HYPODERMIC SINGLE LUMEN NEEDLE FMI BD MEDICAL (BD WEST) MEDICAL SURGICAL 305110 0022962 30382903051107

Patients

Seq Age Sex Outcome Treatment
1 Unknown