640G INSULIN PUMP MMT-1712K
Report
- Report Number
- 2032227-2021-235819
- Event Type
- Malfunction
- Date Received
- December 28, 2021
- Date of Event
- December 24, 2021
- Report Date
- August 2, 2022
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- UDI-DI
- 000000763000317157
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IS
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
(B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE 640G INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.
RETAINER RING = BLACK. CUSTOMER RETURNED PUMP FOR AN ALLEGED BLANK DISPLAY FOUND ON 24-DEC-2022. PUMP PASSES DISPLACEMENT TEST, SLEEP CURRENT TEST, ACTIVE CURRENT TEST AND SELF TEST. NO BLANK DISPLAY NOTED DURING TESTING. SUCCESSFULLY DOWNLOADED HISTORY FILES AND TRACES USING THUS. THE FOLLOWING POWER ALARMS/ALERTS WERE NOTED ON EVENT DATE: LOW BATTERY ALERT [104] ON 12/23/2021 AT START TIME 06:11:00.000, REPLACE BATTERY ALERT [73] ON 12/23/2021 AT START TIME 15:42:00.000, REPLACE BATTERY NOW ALARM [11] ON 12/23/2021 AT START TIME 16:13:00.000, AND POWER LOSS ALARM [6] ON 12/23/2021 AT START TIME 16:27:13.000. THE POWER MANAGEMENT GRAPH CONFIRMED THE UNLOADED VOLTAGE (UL VLITH) AND LOADED VOLTAGE (LOADED VLITH) WERE WITHIN SPEC RANGE. TEST P-CAP AND RESERVOIR LOCKED PROPERLY INTO RESERVOIR COMPARTMENT DURING TESTING. PUMP WAS CUT OPEN TO PERFORM VISUAL INSPECTION AND FOUND MOISTURE DAMAGE INSIDE THE BATTERY TUBE AND MOTOR. TEST P-CAP AND RESERVOIR LOCKED PROPERLY INTO RESERVOIR COMPARTMENT DURING TESTING. THE FOLLOWING WERE NOTED DURING VISUAL INSPECTION: PILLOWING KEYPAD OVERLAY AND PARTIALLY FADED SERIAL NUMBER LABEL. BLANK DISPLAY WAS NOT OBSERVED DURING ANALYSIS, BLANK DISPLAY NOT CONFIRMED. ADDITIONALLY MOISTURE DAMAGE WAS NOTED. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD A BLANK DISPLAY. CUSTOMER STATED THAT THE DISPLAY WAS NOT BLANK LESS THAN 30 SECONDS AND DID NOT RETURN. CUSTOMER STATED THAT THE CONTACTS ON THE BATTERY CAP WERE NOT MISSING OR DAMAGED. CUSTOMER STATED THAT THE BATTERY COMPARTMENT WAS NEITHER DAMAGED NOR CORRODED ALSO THE SPRING WAS NEITHER DAMAGED NOR CORRODED. THE DISPLAY DID NOT RETURN AFTER INSULIN PUMP RESTART. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE DEVICE WILL BE RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2000208 | 640G INSULIN PUMP MMT-1712K | PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-1712K | HG4WPR7 | 000000763000317157 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |