FDA Adverse Event Malfunction Summary report: N

BD FACS¿ SAMPLE PREP ASSISTANT III

MDR report key: 13090577 · Received December 27, 2021

Report

Report Number
2916837-2021-00502
Event Type
Malfunction
Date Received
December 27, 2021
Date of Event
December 10, 2021
Report Date
February 18, 2022
Manufacturer
BECTON, DICKINSON AND COMPANY, BD BIOSCIENCES
Product Code
PER
UDI-DI
00382906472052
PMA / PMN Number
K131301
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H6: INVESTIGATION SUMMARY: SCOPE OF ISSUE: THE SCOPE OF ISSUE IS LIMITED TO PART: 647205 SPAIII AND SERIAL NUMBER: (B)(6). PROBLEM STATEMENT: CUSTOMER REPORTED: SPA IS LEAKING. MANUFACTURING DEFECT TREND: THERE ARE 0 QNS RELATED TO THE REPORTED ISSUE. DATE RANGE (DATE OF INCIDENT TO 12 MONTHS BACK) FROM (B)(6) 2020 TO DATE (B)(6) 2021 (ROLLING 12 MONTHS). COMPLAINT TREND: THERE ARE 11 COMPLAINTS RELATED TO THE REPORTED COMPLAINT. DATE RANGE (DATE OF INCIDENT TO 12 MONTHS BACK) FROM (B)(6) 2020 TO DATE (B)(6) 2021 (ROLLING 12 MONTHS). INVESTIGATION RESULT / ANALYSIS: PER FSE REPORT: REPLACED FLUIDIC TOWER WASTE COUPLING WITH PART ONSITE. REPLACED WASTE FITTING GOING TO WASTE TANK WITH PART ONSITE. TESTED AND VERIFIED INSTRUMENT. SERVICE MAX REVIEW: REVIEW OF RELATED WORK ORDER# (B)(4). INSTALL DATE: 08FEB2018. DEFECTIVE PART NUMBER: DAMAGED COUPLING (NO PART NUMBER PROVIDED). WORK ORDER NOTES: SUBJECT / REPORTED: FLUID LEAK. PROBLEM DESCRIPTION: FLUIDIC LEAK. CAUSE: DAMAGED COUPLING. WORK PERFORMED: REPLACED DAMAGED COUPLING. SOLUTION: REPLACED DAMAGED COUPLING. RETURNED SAMPLE EVALUATION: DID NOT REQUEST RETURN OF DEFECTIVE PARTS. MANUFACTURING DEVICE HISTORY RECORD (DHR) REVIEW: REVIEW OF THE DHR FOR SERIAL NUMBER: (B)(6) WAS REVIEWED. THE INSTRUMENT MET ALL THE MANUFACTURING SPECIFICATIONS PRIOR TO RELEASE. RISK ANALYSIS: RISK MANAGEMENT FILE PART #100245RA, REVISION 03 WAS REVIEWED. HAZARD(S) IDENTIFIED? YES/NO. HAZARD ID: 3.1.29. HAZARD: ENVIRONMENTAL BIOHAZARD. SEVERITY: 5. PROBABILITY: 1. RISK INDEX: 5. IMPLEMENTATION: BD FACS SAMPLE PREP USER¿S GUIDE. RISK CONTROL:ALARP. MITIGATION(S) SUFFICIENT: YES/NO. ROOT CAUSE: BASED ON THE INVESTIGATION RESULT AND THE FSE¿S REPORT THE ROOT CAUSE WAS A DAMAGED COUPLING. CONCLUSION: BASED ON THE INVESTIGATION RESULTS AND THE FSE¿S REPORT THE COMPLAINT WAS CONFIRMED FOR THE SPA LEAK. H3 OTHER TEXT : SEE H10.

Additional Manufacturer Narrative · 0

COMMON DEVICE NAME: AUTOMATED PIPETTING, DILUTING AND SPECIMEN PROCESSING WORKSTATIONS FOR FLOW CYTOMETRIC ANALYSIS. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD FACS¿ SAMPLE PREP ASSISTANT III LEAKED BIOHAZARD THAT WAS NOT CONTAINED WITHIN THE INSTRUMENT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "1) WAS THE LEAK LIQUID OR AIR? LIQUID. 2) WAS THE LEAK CONTAINED WITHIN THE INSTRUMENT? NOT CONTAINED. 3) WAS THERE SPRAY OF LIQUID UNDER PRESSURE? NO. 4) WHAT WAS THE FLUID THAT LEAKED? BIOHAZARD. 5) DID BIOHAZARD LEAK BEFORE OR AFTER WASTE LINE? AFTER WASTE LINE. 6) WAS THE WASTE MIXED WITH DECONTAMINATION/BLEACH? NO. 7) WAS THE CUSTOMER/BD PERSONNEL PHYSICALLY IN CONTACT WITH THE FLUID? NO. IT WAS REPORTED THAT THE SPA IS LEAKING.".

Description of Event or Problem · 0

IT WAS REPORTED THAT BD FACS¿ SAMPLE PREP ASSISTANT III LEAKED BIOHAZARD THAT WAS NOT CONTAINED WITHIN THE INSTRUMENT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: " WAS THE LEAK LIQUID OR AIR? LIQUID. WAS THE LEAK CONTAINED WITHIN THE INSTRUMENT? NOT CONTAINED. WAS THERE SPRAY OF LIQUID UNDER PRESSURE? NO. WHAT WAS THE FLUID THAT LEAKED? BIOHAZARD. DID BIOHAZARD LEAK BEFORE OR AFTER WASTE LINE? AFTER WASTE LINE. WAS THE WASTE MIXED WITH DECONTAMINATION/BLEACH? NO. WAS THE CUSTOMER/BD PERSONNEL PHYSICALLY IN CONTACT WITH THE FLUID? NO. IT WAS REPORTED THAT THE SPA IS LEAKING."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1985792 BD FACS¿ SAMPLE PREP ASSISTANT III SEE H.10 PER BECTON, DICKINSON AND COMPANY, BD BIOSCIENCES 647205 NA 00382906472052

Patients

Seq Age Sex Outcome Treatment
1 Unknown