FDA Adverse Event Malfunction Summary report: N

PENTAX

MDR report key: 13086426 · Received December 27, 2021

Report

Report Number
9610877-2021-50996
Event Type
Malfunction
Date Received
December 27, 2021
Date of Event
December 20, 2021
Report Date
December 27, 2021
Manufacturer
HOYA CORPORATION PENTAX TOKYO OFFICE
Product Code
FDF
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS DEVICE IS CLASSIFIED AS IMPORT FOR EXPORT, THEREFORE 510K IS NOT APPLICABLE. MODEL EC38-I10L-US IS AVAILABLE IN THE USA WITH A 510K NUMBER K131855. WE CHECKED THE RETURNED UNIT AND CONFIRMED THE CCD DRIVER PCB AS DEFECTIVE. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE EXCESSIVE FORCE APPLIED ON THE CCD DRIVER PCB. IN ADDITION, WE CONFIRMED THAT THE ANGLE WIRE PLAYS AND THE ELECTRICAL CONNECTOR CORRODED; HOWEVER, THEY ARE NOT THE MAIN CAUSE, AND/OR IRRELEVANT TO THE ALLEGED COMPLAINT. BASED ON THE TECHNICAL REPORT AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.

Additional Manufacturer Narrative · 0

THIS DEVICE IS CLASSIFIED AS IMPORT FOR EXPORT, THEREFORE 510K IS NOT APPLICABLE. MODEL EC38-I10L-US IS AVAILABLE IN THE USA WITH A 510K NUMBER K131855. WE CHECKED THE RETURNED UNIT AND CONFIRMED THE CCD DRIVER PCB AS DEFECTIVE. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE EXCESSIVE FORCE APPLIED ON THE CCD DRIVER PCB. IN ADDITION, WE CONFIRMED THAT THE ANGLE WIRE PLAYS AND THE ELECTRICAL CONNECTOR CORRODED; HOWEVER, THEY ARE NOT THE MAIN CAUSE, AND/OR IRRELEVANT TO THE ALLEGED COMPLAINT. BASED ON THE TECHNICAL REPORT AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.

Description of Event or Problem · 0

THE TIME OF EVENT IS UNKNOWN. WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE.

Description of Event or Problem · 0

THE TIME OF EVENT IS UNKNOWN. WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1986682 PENTAX VIDEO COLONOSCOPE FDF HOYA CORPORATION PENTAX TOKYO OFFICE EC38-I10NL

Patients

Seq Age Sex Outcome Treatment
1 Unknown