FDA Adverse Event Malfunction Summary report: N

DEXCOM G6 GLUCOSE PROGRAM CONTINUOUS GLUCOSE MONITORING (CGM) SYSTEM

MDR report key: 13077893 · Received December 23, 2021

Report

Report Number
3004753838-2021-433174
Event Type
Malfunction
Date Received
December 23, 2021
Date of Event
December 4, 2021
Report Date
December 23, 2021
Manufacturer
DEXCOM, INC.
Product Code
QDK
PMA / PMN Number
K203089
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

(B)(4).

Description of Event or Problem · 0

IT WAS REPORTED THAT AN INACCURACY BETWEEN THE CONTINUOUS GLUCOSE MONITOR (CGM) AND THE BLOOD GLUCOSE (BG) METER OCCURRED. THE SENSOR WAS INSERTED INTO THE ABDOMEN ON (B)(6) 2021. NO DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. THE REPORTED GLUCOSE VALUES FALL WITHIN THE D ZONE OF THE PARKES ERROR GRID. . NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1980003 DEXCOM G6 GLUCOSE PROGRAM CONTINUOUS GLUCOSE MONITORING (CGM) SYSTEM CONTINUOUS GLUCOSE MONITOR QDK DEXCOM, INC. NI

Patients

Seq Age Sex Outcome Treatment
1 73 YR Male