FDA Adverse Event Malfunction Summary report: N

LACTATE DEHYDROGENASE (LDH)

MDR report key: 13064148 · Received December 22, 2021

Report

Report Number
3002809144-2021-00716
Event Type
Malfunction
Date Received
December 22, 2021
Date of Event
December 14, 2021
Report Date
March 9, 2022
Manufacturer
ABBOTT GMBH
Product Code
CFJ
UDI-DI
00380740161460
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE COMPLAINT INVESTIGATION FOR A FALSELY ELEVATED LACTATE DEHYDROGENASE (LDH) RESULT INCLUDED A SEARCH FOR SIMILAR COMPLAINTS, THE REVIEW OF COMPLAINT TEXT, TRENDING DATA, LABELING, AND DEVICE HISTORY RECORDS. TRENDING REVIEW DETERMINED NO RELATED TRENDS FOR THE PRODUCT. RETURN TESTING WAS NOT COMPLETED AS RETURNS WERE NOT AVAILABLE. DEVICE HISTORY RECORD REVIEW DID NOT IDENTIFY ANY NON-CONFORMANCES OR DEVIATIONS WITH THE LIKELY CAUSE LIST NUMBER AND COMPLAINT ISSUE. MANUFACTURING DOCUMENTATION FOR THE LIKELY CAUSE LIST NUMBER WAS REVIEWED AND DID NOT IDENTIFY ANY ISSUES. LABELING WAS REVIEWED AND SUFFICIENTLY ADDRESSES THE CUSTOMER'S ISSUE. BASED ON THE INFORMATION PROVIDED AND ABBOTT DIAGNOSTICS¿ COMPLAINT INVESTIGATION, NO SYSTEMIC ISSUE OR DEFICIENCY OF LACTATE DEHYDROGENASE (LDH) ASSAY, LOT NUMBER 26340UN21, WAS IDENTIFIED.SECTION G1 - CONTACT OFFICE INFORMATION UPDATED.

Additional Manufacturer Narrative · 0

AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. ALL AVAILABLE PATIENT INFORMATION WAS INCLUDED. ADDITIONAL PATIENT DETAILS ARE NOT AVAILABLE.

Description of Event or Problem · 0

THE CUSTOMER OBSERVED A FALSELY ELEVATED LACTATE DEHYDROGENASE RESULT FOR ONE PATIENT ON AN ARCHITECT C16000 ANALYZER. THE FOLLOWING DATA WAS PROVIDED: INITIAL RESULT WAS 346.64, REPEATS ON OTHER ANALYZERS WERE 141.08, 142.67, AND 143.09 U/L. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1963600 LACTATE DEHYDROGENASE (LDH) NAD REDUCTION/NADH OXIDATION, LACTATE DEHYDROGENASE CFJ ABBOTT GMBH 2P56-22 26340UN21 00380740161460

Patients

Seq Age Sex Outcome Treatment
1 Unknown ARC C16K PRC MOD, 03L77-01, (B)(6)| ARC C16K PRC MOD, 03L77-01,(B)(6)