LIGHT ADJUSTABLE LENS (LAL)
Report
- Report Number
- 3012712027-2021-00087
- Event Type
- Injury
- Date Received
- December 21, 2021
- Date of Event
- November 4, 2021
- Report Date
- December 17, 2021
- Manufacturer
- RXSIGHT, INC.
- Product Code
- PZK
- UDI-DI
- 00818806020203
- PMA / PMN Number
- P160055
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
SITE REPORTED A LIGHT ADJUSTABLE LENS WAS EXPLANTED AS THE LENS WAS NOTED TO BE DISLOCATED, PREVENTING DESIRED REFRACTION. RXSIGHT'S FIRST AWARENESS OF THE EVENT WAS ON (B)(6) 2021. THE DEVICE HISTORY RECORD FOR THE LENS WAS REVIEWED. NO ISSUES WERE NOTED. THE LENS IS IN THE PROCESS OF BEING RETURNED FOR EVALUATION.
SITE REPORTED A LIGHT ADJUSTABLE LENS WAS EXPLANTED AS THE LENS WAS NOTED TO BE DISLOCATED, PREVENTING DESIRED REFRACTION. RXSIGHT'S FIRST AWARENESS OF THE EVENT WAS 12/1/2021. THE EXPLANTED LENS WAS RETURNED FOR EVALUATION. VISUAL INSPECTION AND OPTICAL TESTING OF THE RETURNED LENS CONFIRMED THE PRESENCE OF AN AMBIENT ZONE ON THE LENS, WHICH IS INDICATIVE OF EXPOSURE OF THE LENS TO AMBIENT UV LIGHT PRIOR TO LOCK IN.
SITE REPORTED A LIGHT ADJUSTABLE LENS WAS EXPLANTED AS THE LENS WAS NOTED TO BE DISLOCATED, PREVENTING DESIRED REFRACTION. RXSIGHT'S FIRST AWARENESS OF THE EVENT WAS ON (B)(6) 2021. THE DEVICE HISTORY RECORD FOR THE LENS WAS REVIEWED. NO ISSUES WERE NOTED. THE LENS IS IN THE PROCESS OF BEING RETURNED FOR EVALUATION.
SITE REPORTED A LIGHT ADJUSTABLE LENS WAS EXPLANTED AS THE LENS WAS NOTED TO BE DISLOCATED, PREVENTING DESIRED REFRACTION. RXSIGHT'S FIRST AWARENESS OF THE EVENT WAS ON (B)(6) 2021.
SITE REPORTED A LIGHT ADJUSTABLE LENS WAS EXPLANTED AS THE LENS WAS NOTED TO BE DISLOCATED, PREVENTING DESIRED REFRACTION. RXSIGHT'S FIRST AWARENESS OF THE EVENT WAS 12/1/2021.
SITE REPORTED A LIGHT ADJUSTABLE LENS WAS EXPLANTED AS THE LENS WAS NOTED TO BE DISLOCATED, PREVENTING DESIRED REFRACTION. RXSIGHT'S FIRST AWARENESS OF THE EVENT WAS ON (B)(6) 2021.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1951582 | LIGHT ADJUSTABLE LENS (LAL) | LIGHT ADJUSTABLE LNES | PZK | RXSIGHT, INC. | 60005 | L02-002015 | 00818806020203 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Female | Required Intervention |