FDA Adverse Event Malfunction Summary report: N

BD SYRINGE¿ WITH NEEDLE

MDR report key: 13048235 · Received December 20, 2021

Report

Report Number
3002682307-2021-00665
Event Type
Malfunction
Date Received
December 20, 2021
Date of Event
October 28, 2021
Report Date
March 3, 2022
Manufacturer
BECTON DICKINSON, S.A.
Product Code
FMF
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED MATERIAL NUMBER 301940 AND LOT NUMBER 1902137. THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT AND ALL QUALITY TESTS WERE FOUND TO BE WITHIN SPECIFICATION. AS SAMPLES WERE UNAVAILABLE FOR RETURN, RETAINED SAMPLES WERE OBTAINED FOR FURTHER EVALUATION BY OUR QUALITY ENGINEER TEAM. THROUGH EXAMINATION OF THE RETAINED SAMPLES, NO SIGNS OF DEFECT COULD BE OBSERVED. BASED ON THE LIMITED INVESTIGATION RESULTS, AN EXACT CAUSE COULD NOT BE DETERMINED FOR THE REPORTED INCIDENT.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHITE FOREIGN MATTER WAS FOUND IN THE BD SYRINGE¿ WITH NEEDLE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM (B)(6): "(B)(6) 2021, WHEN THE NURSE WAS DISPENSING MEDICINE FOR THE PATIENT, SHE TOOK A 2ML SYRINGE AND FOUND A WHITE LINEAR FOREIGN MATTER IN THE SYRINGE, WHICH WAS REPLACED."

Description of Event or Problem · 0

IT WAS REPORTED THAT WHITE FOREIGN MATTER WAS FOUND IN THE BD SYRINGE¿ WITH NEEDLE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM CHINESE: "2021-10-28, WHEN THE NURSE WAS DISPENSING MEDICINE FOR THE PATIENT, SHE TOOK A 2ML SYRINGE AND FOUND A WHITE LINEAR FOREIGN MATTER IN THE SYRINGE, WHICH WAS REPLACED."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1945147 BD SYRINGE¿ WITH NEEDLE PISTON SYRINGE FMF BECTON DICKINSON, S.A. 1902137

Patients

Seq Age Sex Outcome Treatment
1 Unknown