BD INTIMA II¿ IV CATHETER WITH PRN ADAPTER
Report
- Report Number
- 3014704491-2021-00364
- Event Type
- Malfunction
- Date Received
- December 20, 2021
- Date of Event
- November 15, 2021
- Report Date
- December 3, 2021
- Manufacturer
- BECTON DICKINSON MEDICAL DEVICES CO., LTD. SUZHOU
- Product Code
- FOZ
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
INVESTIGATION SUMMARY: IN RESPONSE TO THE EVENT REPORTED A DEVICE HISTORY REVIEW WAS CONDUCTED FOR LOT NUMBER 0231208. OUR RECORDS SHOW THAT THIS IS THE ONLY INSTANCE OF THIS ISSUE OCCURRING IN THIS PRODUCTION BATCH. ACCORDING TO THE SAMPLING PLAN APPLIED FOR PRODUCT PERFORMANCE, THIS LOT WAS ACCEPTED AND RELEASED WITHOUT DEFECTS BEING NOTED DURING THE FINAL ASSEMBLY OR VISUAL INSPECTIONS. A SAMPLE COULD NOT BE OBTAINED FOR EVALUATION AND TESTING; IN LIEU OF THE AFFECTED DEVICE, FUNCTIONAL TESTING WAS PERFORMED ON RETENTION SAMPLES FOR THIS LOT, THE RESULTS OF THESE SHOW THAT THE TESTED UNITS PERFORMED WITHIN PRODUCT SPECIFICATIONS. UNFORTUNATELY WITHOUT THE ABILITY TO INVESTIGATE THE AFFECTED UNIT OUR QUALITY ENGINEERS WERE UNABLE TO DETERMINE THE ROOT CAUSE FOR THIS COMPLAINT.
IT WAS REPORTED WHEN USING THE BD INTIMA II¿ IV CATHETER WITH PRN ADAPTER, THE DEVICE EXPERIENCED LEAKAGE AT THE CATHETER AND HUB JUNCTION. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER. THE CUSTOMER STATED: AFTER THE PUNCTURE WAS SUCCESSFUL, THE JOINT WAS TIGHTENED AND THE PATIENT'S SHEETS WERE FOUND TO BE WET AT 9:00. LATER, IT WAS FOUND THAT THERE WAS FLUID LEAKAGE AT THE JOINT OF THE PATIENT'S VENOUS INDWELLING NEEDLE, AND THE INDWELLING NEEDLE WAS IMMEDIATELY REPLACED WITH A RE-PUNCTURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1945886 | BD INTIMA II¿ IV CATHETER WITH PRN ADAPTER | INTRAVASCULAR CATHETER | FOZ | BECTON DICKINSON MEDICAL DEVICES CO., LTD. SUZHOU | 0231208 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |