HD 3CMOS AUTOCLAVABLE CAMERA HEAD

MDR report key: 13041098 · Received December 19, 2021

Report

Report Number
8010047-2021-16396
Event Type
Malfunction
Date Received
December 19, 2021
Date of Event
November 23, 2021
Report Date
February 15, 2022
Manufacturer
OLYMPUS MEDICAL SYSTEMS CORP.
Product Code
OWN
PMA / PMN Number
K200542
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE ADDITIONAL INFORMATION. OMSC REVIEWED THE MANUFACTURING HISTORY (DHR) OF THE SUBJECT DEVICE AND CONFIRMED NO IRREGULARITY. <E226 > A CABLE FAILURE HAD BEEN CONFIRMED DURING EVALUATION. THEREFORE, IT IS POSSIBLE THAT E226 OCCURRED BECAUSE THE VIDEO FUNCTION DID NOT OPERATE NORMALLY DUE TO A CABLE FAILURE. <NO IMAGE> CCD FAILURE HAD BEEN CONFIRMED DURING EVALUATION. THEREFORE, IT IS POSSIBLE THAT NO IMAGE OCCURRED BECAUSE THE VIDEO FUNCTION DID NOT OPERATE NORMALLY DUE TO THE FAILURE OF THE CCD. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, THIS REPORT WILL BE SUPPLEMENTED.

Additional Manufacturer Narrative · 0

THE DEVICE WAS INSPECTED AT OAZ. OAZ CHECKED THE SUBJECT DEVICE AND FOUND THE REPORTED PHENOMENA AND THE SCOPE COMMUNICATION ERROR E226 WAS CAUSED BY THE FAILURE OF THE CABLE AND "NO IMAGE" WAS CAUSED BY THE FAILURE OF THE CCD. THE EXACT CAUSE OF THE REPORTED EVENT COULD NOT BE CONCLUSIVELY DETERMINED AT THIS TIME. IF ADDITIONAL INFORMATION IS RECEIVED, THIS REPORT WILL BE SUPPLEMENTED.

Description of Event or Problem · 0

OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) WAS INFORMED THAT DURING THE INCOMING INSPECTION FOR REPAIR AT THE SERVICE DEPARTMENT OF OLYMPUS AUSTRALIA (OAZ), IT WAS FOUND THAT THE SCOPE COMMUNICATION ERROR E226 OCCURRED AND THE ENDOSCOPIC IMAGE WAS NOT DISPLAYED. THE SUBJECT DEVICE HAD BEEN RETURNED TO OAZ FOR REPAIR OF THE FLICKERING IMAGE. THERE WAS NO REPORT OF PATIENT INJURY ASSOCIATED WITH THE EVENT.