BD VACUTAINER® EST Z (NO ADDITIVE) PLUS BLOOD COLLECTION TUBES
Report
- Report Number
- 9617032-2021-01299
- Event Type
- Malfunction
- Date Received
- December 17, 2021
- Date of Event
- November 16, 2021
- Report Date
- November 29, 2021
- Manufacturer
- BECTON, DICKINSON AND COMPANY (BD)
- Product Code
- FMH
- UDI-DI
- 50382903627252
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AU
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
INVESTIGATION SUMMARY: BD HAD NOT RECEIVED SAMPLES, BUT FIVE (5) PHOTOS WERE PROVIDED FOR INVESTIGATION. THE PHOTOS WERE REVIEWED AND THE INDICATED FAILURE MODE FOR STOPPER CORING WITH THE INCIDENT LOT WAS NOT OBSERVED. THREE (3) PHOTOS SHOW THE TESTING EQUIPMENT AND THE OTHER TWO (2 ) PHOTOS SHOW UNUSED TUBES. ADDITIONALLY, TEN (10) RETENTION SAMPLES FROM BD INVENTORY WERE TAKEN, DRAWN WITH WATER AND NO ISSUES WERE OBSERVED RELATING TO STOPPER CORING AS ALL SAMPLES MET SPECIFICATIONS. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY NOTIFICATIONS. THIS COMPLAINT IS UNABLE TO BE CONFIRMED FOR THE INDICATED FAILURE MODE OF STOPPER CORING. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.
IT WAS REPORTED WHEN USING THE BD VACUTAINER® EST Z (NO ADDITIVE) PLUS BLOOD COLLECTION TUBES, THE DEVICE EXPERIENCED A CLOGGED/BLOCKED INSTRUMENTATION PROBE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER. THE CUSTOMER STATED: SPA III QUICK CONNECTOR TO THE FILTER WASTE FILTER IS CLOGGED BECAUSE OF THE STOPPER OF THE TUBES. THE BD SYSTEM SPA III HAS TO CONSTANTLY BE WORKED ON AND UNCLOGGED BY OUR TECHNICAL SERVICE BECAUSE THE QUICK CONNECTOR TO THE WASTE FILTER CLOGS UP BECAUSE OF THE STOPPER OF THE TUBES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1922202 | BD VACUTAINER® EST Z (NO ADDITIVE) PLUS BLOOD COLLECTION TUBES | SPECIMEN TRANSPORT AND STORAGE CONTAINER. | FMH | BECTON, DICKINSON AND COMPANY (BD) | 362725 | 1085023 | 50382903627252 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |