FDA Adverse Event Injury Summary report: N

ISOLETTE 8000

MDR report key: 13023270 · Received December 16, 2021

Report

Report Number
2510954-2021-00004
Event Type
Injury
Date Received
December 16, 2021
Date of Event
August 21, 2021
Report Date
December 16, 2021
Manufacturer
DRAEGER MEDICAL INFANT CARE, INC.
Product Code
FMZ
PMA / PMN Number
K001242
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
RISK MANAGER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION WAS PROVIDED NOTING THE PATIENT WAS TRANSFERRED TO A DIFFERENT HOSPITAL FOR EVALUATION AND VIDEO IMAGING. THE DEPRESSION FRACTURE AND SUBDUAL HEMATOMA WERE CONFIRMED. THE BABY WAS DISCHARGED THE FOLLOWING DAY. THE HOSPITAL WAS TO DO A FOLLOW UP APPOINT IN A MONTH. PATIENT SEEMS TO BE RECOVERING. THE FOLLOWING FURTHER DETAILS WERE PROVIDED REGARDING THE EVENT. THE NURSE WENT TO CHECK AND FEED THE BABY. SHE OPENED BOTH PORT HOLES AND PUT ARMS IN TO FEED THE BABY, BED WAS FLAT. SHE HAND RAISED THE BABY TO FEED. THE NURSE DID STATE THAT SHE BELIEVES SHE CLOSED THE PORT HOLE DOOR, HOWEVER WHEN THE BABY WAS FOUND THE RIGHT PORT HOLD DOOR WAS OPEN AND THE BABY WAS FOUND BETWEEN THE WALL AND THE I8000 ON THE FLOOR. THE I8000 WAS EVALUATED ON SITE BY THE HOSPITAL BIOMED, THE HEAD OF CLINICAL ENGINEERING AND THE RISK MANAGER. THE LATCHES WERE TESTED AND PUSHED AND PULLED ON TO CONFIRM THEY LATCH PROPERLY AND WERE NOT WEAKENED. THE SITE INVESTIGATION IDENTIFIED ROOT CAUSE AS HUMAN ERROR AND CONFIRMED THAT THE UNIT DID NOT PLAY A FACTOR IN THE PATIENT OUTCOME. THERE WAS NO DRAEGER DEVICE MALFUNCTION. THE INSTRUCTIONS FOR USE INCLUDES AN OPERATIONAL CHECKOUT SECTION WHICH INCLUDES A HOOD/SHELL OPERATIONAL CHECKOUT PROCEDURE THAT INSTRUCTS THE USER TO CHECK THE HAND PORT LATCHES FOR PROPER OPERATION BY ¿PRESSING THE LATCH ON EACH HAND PORT AND VERIFY THAT DOOR SWINGS OPEN¿, AND ¿CLOSE THE DOORS AND CHECK FOR PROPER LATCHING AND QUIETNESS.¿ WARNING - RISK OF DEATH OR SERIOUS INJURY. POSITIVELY SECURE ALL HAND PORT LATCHES AND ACCESS PANEL LOCKING KNOBS TO AVOID ACCIDENTAL OPENING. WARNING - RISK OF DEATH OR SERIOUS INJURY. FOR INFANT SAFETY, DO NOT LEAVE THE INFANT UNATTENDED WHEN THE ACCESS PANELS OR HAND PORTS ARE OPEN. NOTE THAT THIS CASE WAS INITIALLY REPORTED UNDER MDR# 1220063-2021-00028 IN ERROR, THE CORRECT MDR# IS 2510954-2021-00004.

Description of Event or Problem · 0

DRÄGER WAS NOTIFIED THROUGH A MEDWATCH REPORT RECEIVED WHERE IS WAS REPORTED THAT A PATIENT HAD FALLEN OUT OF AN ISOLETTE 8000 AND SUSTAINED A HEAD DEPRESSION FRACTURE AND SUBDURAL HEMATOMA. ADDITIONALLY THE UNIT WAS INSPECTED BY THE BIOMEDICAL DEPARTMENT AND CONFIRMED THAT THE DOOR LATCHING MECHANISMS WERE INTACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1921252 ISOLETTE 8000 INFANT INCUBATOR WITH OXYGEN, SCALE AND HUMDITY FMZ DRAEGER MEDICAL INFANT CARE, INC. MU21124

Patients

Seq Age Sex Outcome Treatment
1 5 DA Female Required Intervention