FDA Adverse Event Injury Summary report: N

NEWTON IQ W/ STAY SAFE CYCLER SET

MDR report key: 1302189 · Received January 30, 2009

Report

Report Number
8030665-2009-00006
Event Type
Injury
Date Received
January 30, 2009
Date of Event
December 31, 2008
Report Date
January 29, 2009
Manufacturer
REYNOSA MANUFACTURING
Product Code
FKX
PMA / PMN Number
K811986
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

A REPORT HAS BEEN RECEIVED FROM A PERITONEAL DIALYSIS UNIT THAT REPORTED A PINHOLE LEAK CAUSED AN EPISODE OF PERITONITIS. THE RN WAS CONTACTED FOR INFORMATION REGARDING THIS EVENT. IT HAS BEEN LEARNED THAT THE PT HAD USED THESE LINES, AND AFTER HE CAME OFF THE CYCLER, HE HAD NOTICED HIS FLOOR WAS WET. HE FOLLOWED THE FLUID PATH AND FOUND THAT THE FLUID HAD COME OUT OF A PINHOLE AT THE PUMP SEGMENT. THE RN STATED THAT ABOUT 6 HOURS LATER, THIS PT DEVELOPED ABDOMEN PAIN AND HAD CLOUDY BAGS (EFFLUENT). THE PT HAD COME INTO THE CLINIC FOR EVALUATION. AT THAT TIME, A SAMPLE OF THE EFFLUENT WAS OBTAINED FOR A CULTURE. THE RESULTS OF THE CULTURE HAD INDICATED NO GROWTH, HOWEVER THE RN SAW THE CLOUDY BAGS AND TREATED THE PT PER PROTOCOL. AS A RESULT, THE PT RECEIVED INTRAPERITONEAL ANTIBIOTICS FOR PERITONITIS FOR THIS EVENT. THE PT HAS SAVED THE SAMPLE AS ADVISED BY THE RN. ADDITIONALLY, IT HAS BEEN LEARNED THAT THE PT IS FINE AND IS ABLE TO CONTINUE PERITONEAL DIALYSIS AS ORDERED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NEWTON IQ W/ STAY SAFE CYCLER SET PERITONEAL DIALYSIS TREATMENT SET FKX REYNOSA MANUFACTURING NA 8LR114

Patients

Seq Age Sex Outcome Treatment
1 NI Other| R