FDA Adverse Event Malfunction Summary report: N

DIGORA OPTIME

MDR report key: 13018602 · Received December 16, 2021

Report

Report Number
3005383085-2021-00001
Event Type
Malfunction
Date Received
December 16, 2021
Date of Event
November 15, 2021
Report Date
December 16, 2021
Manufacturer
PALODEX GROUP OY
Product Code
EHD
PMA / PMN Number
K041050
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS ISSUE RELATES TO THE 7-8 YEARS OLD DEFECTIVE POWER SUPPLY UNIT (PSUS) OF DIGORA OPTIME / SCAN EXAM SCANNERS. FDA IS AWARE OF THE FIELD ACTION AND IT IS COVERED IN RECALLS Z-0839-2020 - Z-0842-2020. THIS IS THE FIRST REPORT WE HAVE RECEIVED OF A USER IN THE UNITED STATES EXPERIENCING THIS MALFUNCTION. THIS CUSTOMER HAD RECEIVED 3 NOTIFICATIONS REGARDING THE FIELD ACTION SENT TO THE DENTAL PRACTICE ON 08/25/2020, 09/30/2021 AND 11/5/2021 AND HAD NOT YET ORDERED A REPLACEMENT PSU. THE DEVICE IS BEING REPLACED.

Additional Manufacturer Narrative · 0

THIS ISSUE RELATES TO THE 7-8 YEARS OLD DEFECTIVE POWER SUPPLY UNIT (PSUS) OF DIGORA OPTIME / SCAN EXAM SCANNERS. FDA IS AWARE OF THE FIELD ACTION AND IT IS COVERED IN RECALLS Z-0839-2020 - Z-0842-2020. THIS IS THE FIRST REPORT WE HAVE RECEIVED OF A USER IN THE UNITED STATES EXPERIENCING THIS MALFUNCTION. THIS CUSTOMER HAD RECEIVED 3 NOTIFICATIONS REGARDING THE FIELD ACTION SENT TO THE DENTAL PRACTICE ON 08/25/2020, 09/30/2021 AND 11/5/2021 AND HAD NOT YET ORDERED A REPLACEMENT PSU. THE DEVICE IS BEING REPLACED.

Description of Event or Problem · 0

DENTAL IMAGING PLATE SCANNER DEVICE'S POWER SUPPLY UNIT HAS HAD A SHORT CIRCUIT. DENTIST STATES THAT THERE WAS NOT A FIRE BUT MORE SO THE POWER SUPPLY SHORTING AND MELTING TO THE INSIDE OF THE UNIT. USER SAYS NO PATIENT WAS INVOLVED, NO INJURY TO PERSONNEL OR PROPERTY DAMAGE OCCURRED. DEVICE IS NOW NOT FUNCTIONAL.

Description of Event or Problem · 0

DENTAL IMAGING PLATE SCANNER DEVICE'S POWER SUPPLY UNIT HAS HAD A SHORT CIRCUIT. DENTIST STATES THAT THERE WAS NOT A FIRE BUT MORE SO THE POWER SUPPLY SHORTING AND MELTING TO THE INSIDE OF THE UNIT. USER SAYS NO PATIENT WAS INVOLVED, NO INJURY TO PERSONNEL OR PROPERTY DAMAGE OCCURRED. DEVICE IS NOW NOT FUNCTIONAL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1917037 DIGORA OPTIME EXTRAORAL SOURCE X-RAY SYSTEM EHD PALODEX GROUP OY DXR-50

Patients

Seq Age Sex Outcome Treatment
1 Unknown