FDA Adverse Event Malfunction Summary report: N

GAP SPACER

MDR report key: 13017781 · Received December 16, 2021

Report

Report Number
1020279-2021-08695
Event Type
Malfunction
Date Received
December 16, 2021
Date of Event
November 29, 2021
Report Date
February 25, 2022
Manufacturer
SMITH & NEPHEW, INC.
Product Code
HWT
UDI-DI
00885556578445
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H3, H6: THE ASSOCIATED DEVICE WAS RETURNED AND EVALUATED. A VISUAL INSPECTION WAS CONDUCTED AND CONFIRMED THE DEVICE IS BROKEN IN HALF, RENDERING THE DEVICE INOPERABLE. A REVIEW OF COMPLAINT HISTORY BASED ON THE HISTORICAL DATA REVEALED SIMILAR EVENTS FOR THE LISTED BATCH, THIS FAILURE MODE WILL BE MONITORED FOR FUTURE COMPLAINTS FOR ANY NECESSARY CORRECTIVE ACTIONS. A REVIEW OF THE MANUFACTURING RECORDS DID NOT REVEAL A MANUFACTURING ABNORMALITY THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED INCIDENT. AT THIS TIME, WE DO NOT HAVE REASON TO SUSPECT THAT THE PRODUCT FAILED TO MEET ANY PRODUCT SPECIFICATIONS AT THE TIME OF MANUFACTURE. A REVIEW OF THE RISK MANAGEMENT FILE REVEALED THIS FAILURE MODE WAS PREVIOUSLY IDENTIFIED. THE ANTICIPATED RISK LEVEL IS STILL ADEQUATE. ASSESSMENT OF HISTORICAL ESCALATED CASES CONCLUDED THAT THERE ARE NO PRIOR ACTIONS RELATED TO THIS DEVICE AND FAILURE MODE. A CONTRIBUTION OF THE DEVICE TO THE REPORTED EVENT COULD BE CORROBORATED AS THE DEVICE SHOWS SIGNS OF DAMAGE/WEAR. THIS DEVICE IS A REUSABLE INSTRUMENT THAT CAN BE EXPOSED TO NUMEROUS SURGERIES. DAMAGE FROM PROLONGED USE, MISUSE OR ROUGH HANDLING ARE PROBABLE CAUSES OF THE REPORTED EVENT. WE RECOMMEND THAT ALL REUSABLE INSTRUMENTS BE ROUTINELY INSPECTED FOR WEAR AND DAMAGE AND REPLACED AS NECESSARY. BASED ON THIS INVESTIGATION, THE NEED FOR CORRECTIVE ACTION IS NOT INDICATED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE COMPLAINT WILL BE REOPENED. NO FURTHER INVESTIGATION IS WARRANTED FOR THIS COMPLAINT; HOWEVER, WE WILL CONTINUE TO MONITOR FUTURE COMPLAINTS AND INVESTIGATE AS NECESSARY. WE CONSIDER THIS INVESTIGATION CLOSED.

Additional Manufacturer Narrative · 0

INTERNAL REFERENCE NUMBER: CASE (B)(4).

Description of Event or Problem · 0

IT WAS REPORTED THAT, DURING TOTAL KNEE ARTHROPLASTY, A JOURNEY GAP SPACER BROKE OUTSIDE THE PATIENT. SURGERY WAS RESUMED, WITHOUT ANY DELAY, WITH A SMITH & NEPHEW BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM. NO PIECES FELL INTO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1916630 GAP SPACER TEMPLATE HWT SMITH & NEPHEW, INC. 71935180 18LDN1935 00885556578445

Patients

Seq Age Sex Outcome Treatment
1 Unknown