IMPL TAPERED SCR-V SBM 4. 7MM 4.5MM 10MM
Report
- Report Number
- 0002023141-2021-03617
- Event Type
- Injury
- Date Received
- December 16, 2021
- Date of Event
- June 26, 2021
- Report Date
- June 8, 2022
- Manufacturer
- ZIMMER DENTAL
- Product Code
- DZE
- UDI-DI
- 00889024020047
- PMA / PMN Number
- K061410
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- DENTIST
- Health Professional
- Yes
Narratives
ZIMMER BIOMET COMPLAINT NUMBER (B)(4). AN IMPL TAPERED SCR-V SBM 4. 7MM 4.5MM 10MM (TSVWB10) AND TOOL IMPLANT REMOVAL SCR TYPE IMP (IRT) WAS RETURNED FOR INVESTIGATION. VISUAL INSPECTION OF THE AS RETURNED PRODUCT IDENTIFIED WORN MARKINGS DUE TO USAGE AND A FRACTURE AT THE HEX OF A MOUNT AND REMOVAL TOOL INSIDE OF IMPLANT. DEVICE HISTORY RECORD (DHR) REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER (1242120). IT WAS CONFIRMED THAT ALL OPERATIONS AND INSPECTIONS WERE EXECUTED AS PER APPLICABLE PROCEDURE. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT WAS NOTED AS PART OF THE DHR. LOT WAS INSPECTED AND PASSED ALL ACCEPTANCE CRITERIA BY QA. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER (1242120) FOR SIMILAR EVENT AND NO OTHER COMPLAINT WAS IDENTIFIED. BASED ON THE AVAILABLE INFORMATION, DEVICE MALFUNCTION DID OCCUR AND THE REPORTED EVENT WAS CONFIRMED.
ZIMMERBIOMET COMPLAINT NUMBER (B)(4). WEIGHT UNKNOWN / NOT PROVIDED. EMAIL ADDRESS UNKNOWN / NOT PROVIDED. PMA/510(K) NUMBER: K011028 AND K013227.
NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.
THE DOCTOR REPORTED THAT THE MOUNT FRACTURED INSIDE THE IMPLANT AND WAS REMOVED. PATIENT WOULD HAVE TO RETURN TO PLACE A NEW IMPLANT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1918323 | IMPL TAPERED SCR-V SBM 4. 7MM 4.5MM 10MM | DENTAL IMPLANT | DZE | ZIMMER DENTAL | TSVWB10 | 1242120 | 00889024020047 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 44 YR | Female | Required Intervention |