FDA Adverse Event Malfunction Summary report: N

PENTAX

MDR report key: 13015056 · Received December 15, 2021

Report

Report Number
9610877-2021-01824
Event Type
Malfunction
Date Received
December 15, 2021
Date of Event
December 3, 2021
Report Date
March 14, 2022
Manufacturer
HOYA CORPORATION PENTAX TOKYO OFFICE
Product Code
FDF
UDI-DI
04961333237869
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

IF ADDITIONAL INFORMATION BECOMES AVAILABLE, A SUPPLEMENTAL REPORT WILL BE FILED WITH THE NEW INFORMATION. THIS DEVICE IS CLASSIFIED AS IMPORT FOR EXPORT, THEREFORE 510K IS NOT APPLICABLE. MODEL EC38-I10CL-US IS AVAILABLE IN THE USA WITH A 510K NUMBER K190805.

Additional Manufacturer Narrative · 0

EVALUATION SUMMARY: THIS DEVICE IS CLASSIFIED AS IMPORT FOR EXPORT, THEREFORE 510K IS NOT APPLICABLE. MODEL EC38-I10CL-US IS AVAILABLE IN THE USA WITH A 510K NUMBER K190805. WE CHECKED THE RETURNED UNIT AND CONFIRMED THE CCD MODULE AS DEFECTIVE. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE EXCESSIVE FORCE APPLIED ON THE CCD MODULE. IN ADDITION, WE CONFIRMED THAT THE OBJECTIVE LENS DIRTY; HOWEVER, IT IS NOT THE MAIN CAUSE, AND/OR IRRELEVANT TO THE ALLEGED COMPLAINT. BASED ON THE TECHNICAL REPORT AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.

Description of Event or Problem · 0

MANY SPOTS IN THE IMAGE, MOISTURE UNDER THE LED . THIS EVENT OCCURRED AT THE TIME OF DURING INSPECTION. THERE WAS NO REPORT OF PATIENT HARM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1911129 PENTAX VIDEO COLONO SCOPE FDF HOYA CORPORATION PENTAX TOKYO OFFICE EC38-I10CL 04961333237869

Patients

Seq Age Sex Outcome Treatment
1 Unknown