V.A.C. THERAPY SYSTEM
Report
- Report Number
- 1625774-2009-00014
- Event Type
- Injury
- Date Received
- January 30, 2009
- Date of Event
- April 24, 2008
- Report Date
- April 24, 2008
- Manufacturer
- KCI USA, INC.
- Product Code
- JCX
- PMA / PMN Number
- K062227
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- NURSE
Narratives
THE PHYSICAL THERAPIST CARING FOR THE PT STATED THAT SHE DID NOT BELIEVE THAT V.A.C. THERAPY PLAYED A ROLE IN THE ALLEGED EVENT. ACCORDING TO THE PHYSICAL THERAPIST, THE PT WAS NON-COMPLIANT WITH THERAPY. THE REPRESENTATIVE STATED THAT THE PT WOULD REMOVE THE V.A.C. DRESSING ON THEIR OWN AND NOT COMMUNICATE TO HIS CAREGIVERS THE NUMBER OF FOAM PIECES THAT WERE REMOVED. THIS APPEARS TO BE A CASE OF USE ERROR (PT NON-COMPLIANCE), WHERE THE FOAM WAS LEFT IN THE PT'S WOUND WHICH RESULTED IN THE FOAM BEING SURGICALLY REMOVED. V.A.C. LABELING WARNS UNDER "FOAM PLACEMENT": "DO NOT PLACE ANY FOAM DRESSING UNTO BLIND/UNEXPLORED TUNNELS. THE V.A.C. WHITEFOAM DRESSING MAY BE MORE APPROPRIATE FOR USE WITH EXPLORED TUNNELS. ALWAYS COUNT THE TOTAL NUMBER OF PIECES OF FOAM USED IN THE WOUND AND DOCUMENT THAT NUMBER ON THE DRAPE AND IN THE PT'S CHART." IT ALSO STATES UNDER "FOAM REMOVAL": V.A.C. FOAM DRESSINGS ARE NOT BIOABSORBABLE. ALWAYS COUNT THE TOTAL NUMBER OF PIECES OF FOAM REMOVED FROM THE WOUND AND ENSURE THE SAME NUMBER OF FOAM PIECES WAS REMOVED AS PLACED. FOAM LEFT IN THE WOUND FOR GREATER THAN THE RECOMMENDED TIME PERIOD MAY FOSTER IN-GROWTH OF TISSUE INTO THE FOAM, CREATE DIFFICULTY IN REMOVING THE FOAM FROM THE WOUND, OR LEAD OF INFECTION OR OTHER ADVERSE EVENTS."
IT WAS REPORTED THAT A MALE PT RECEIVING V.A.C. THERAPY FOR BILATERAL ISCHIAL PRESSURE ULCERS REQUIRED TWO SURGERIES FOR REMOVAL OF PIECES OF V.A.C. GRANUFOAM DRESSING THAT WERE INADVERTENTLY LEFT IN A TUNNEL OF THE PT'S WOUND. ACCORDING TO THE PHYSICAL THERAPIST CARING FOR THE PT, A PIECE OF V.A.C. GRANUFOAM WAS REMOVED BY SHARP DEBRIDEMENT DURING A SURGICAL PROCEDURE THAT WAS PERFORMED IN 2008. THE PHYSICAL THERAPIST STATED THAT V.A.C. THERAPY WAS THEN REAPPLIED. DURING THE PT'S NEXT VISIT THE FOLLOWING MONTH, ANOTHER PIECE OF FOAM, BELIEVED TO HAVE BEEN RETAINED FROM THE ORIGINAL EPISODE, WAS FOUND IN THE PT'S WOUND. THE PIECE OF FOAM WAS AGAIN REMOVED BY SHARP DEBRIDEMENT AND V.A.C. THERAPY WAS REAPPLIED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | V.A.C. THERAPY SYSTEM | ACCESSORY TO JCX | JCX | KCI USA, INC. | V.A.C. GRANUFOAM | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 21 YR | Required Intervention |