FDA Adverse Event Injury Summary report: N

SMR GLENOSPHERE Ø 40MM

MDR report key: 13010706 · Received December 15, 2021

Report

Report Number
3008021110-2021-00101
Event Type
Injury
Date Received
December 15, 2021
Date of Event
April 10, 2020
Report Date
December 10, 2021
Manufacturer
LIMACORPORATE S.P.A.
Product Code
KWS
PMA / PMN Number
K163397
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

BY CHECKING THE MANUFACTURING CHARTS OF THE INVOLVED LOT #1922985, NO PRE-EXISTING ANOMALY WAS FOUND ON A TOTAL OF 37 ITEMS MANUFACTURED WITH THE SAME LOT #. ACCORDING TO OUR RECORDS, AT LEAST 36 OUT OF 37 GLENOSPHERES WITH LOT #1922985 - STER. 2000011 HAVE BEEN IMPLANTED AND THIS IS THE ONLY COMPLAINT RECEIVED ON THIS LOT #. EXPLANTED ITEMS WERE NOT AVAILABLE TO BE RETURNED TO LIMACORPORATE FOR FURTHER ANALYSIS. NO ADDITIONAL DETAILS WERE AVAILABLE ON THIS POST-OPERATIVE ISSUE, SPECIFICALLY PRE-OPERATIVE X-RAYS RELATED TO THE REVISION SURGERY WERE REQUESTED TO THE COMPLAINT SOURCE, BUT THEY WERE NOT AVAILABLE. BASED ON THE FEW INFORMATION RECEIVED, WE ARE NOT ABLE TO FURTHER INVESTIGATE THE ROOT CAUSE OF THE EVENT. HOWEVER, CONSIDERING THAT THE CHECK OF THE MANUFACTURING CHARTS HIGHLIGHTED NO ANOMALIES ON THE TOTAL NUMBER OF COMPONENTS MANUFACTURED WITH LOT #1922985, WE CAN STATE THAT THE EVENT WAS NOT PRODUCT RELATED. PMS DATA ACCORDING TO LIMACORPORATE PMS DATA, THE REVISION RATE OF SMR REVERSE IMPLANTS DUE TO DISLOCATION IS 0.15%. BASED ON THE ROOT CAUSE ANALYSIS PERFORMED AND ACCORDING TO THE RELEVANT PMS DATA, NO CORRECTIVE ACTIONS REQUIRED FOR THIS SPECIFIC CASE. LIMACORPORATE WILL CONTINUE MONITORING THE MARKET TO PROMPTLY DETECT ANY FURTHER SIMILAR ISSUE. NOTE: THIS IS A COMBINED INITIAL-FINAL MDR.

Description of Event or Problem · 0

SHOULDER REVISION SURGERY OF A SMR REVERSE PROSTHESIS PERFORMED ON (B)(6) 2020, DUE TO IMPLANT DISLOCATION. IT WAS REPORTED THAT THE PATIENT HAD MILD SHOULDER PAIN AND UPON EXAMINATION, IT WAS NOTED THAT THE PROSTHESIS WAS DISLOCATED. THE FOLLOWING COMPONENTS WERE EXPLANTED AND REPLACED: SMR GLENOSPHERE Ø 40MM (PRODUCT CODE 1374.09.121, LOT #1922985 - STER. 2000011), SMR REVERSE LINER STD D.40MM (PRODUCT CODE 1365.50.810, LOT #19AT2TJ - STER. 2000002), SMR REVERSE HUMERAL BODY (PRODUCT CODE 1352.15.010, LOT #1921258 - STER. 2000016), AND SMR SMALL-R CONNECTOR +2 (PRODUCT CODE 1374.15.312, LOT #1914234 - STER. 1900322). A +4MM CONNECTOR AND SHORT REVERSE HUMERAL BODY WERE IMPLANTED. PATIENT'S HISTORY OF REVISION SURGERIES IS THE FOLLOWING: SMR REVERSE PROSTHESIS IMPLANTED ON (B)(6) 2020; FIRST REVISION SURGERY PERFORMED ON (B)(6) 2020 DUE TO IMPLANT DISLOCATION. THE EVENT WAS REGISTERED AS COMPLAINT #(B)(4) AND REPORTED TO THE FDA BY MFR #3008021110-2021-00100); SECOND REVISION SURGERY PERFORMED ON (B)(6) 2020, DUE TO IMPLANT DISLOCATION (EVENT REPORTED ON THIS FORM); THIRD REVISION SURGERY PERFORMED ON (B)(6) 2020, DUE TO INFECTION. A CEMENT SPACER WAS IMPLANTED. THE EVENT WAS REGISTERED AS COMPLAINT #(B)(4) AND REPORTED TO THE FDA BY MFR #3008021110-2021-00102. IT WAS FURTHER REPORTED THAT THE CEMENT SPACER ALSO DISLOCATED POST (B)(6) SURGERY. A CUSTOMIZED SHOULDER IMPLANT WAS REQUESTED. PATIENT IS A MALE, (B)(6) . IT WAS REPORTED HE HAS TYPE 2 DIABETES. HE SUFFERS OF GASTROESOPHAGEAL REFLUX DISEASE (GERD), HYPERCHOLESTEREMIA AND HYPERTENSION. EVENT HAPPENED IN THE US.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1911844 SMR GLENOSPHERE Ø 40MM GLENOSPHERE DIA. 40 MM KWS LIMACORPORATE S.P.A. 1374.09.121 1922985

Patients

Seq Age Sex Outcome Treatment
1 Unknown Other