FDA Adverse Event Malfunction Summary report: N

BROVIAC CV CATHETERS (WHITE ADAPTER)

MDR report key: 13009721 · Received December 15, 2021

Report

Report Number
13009721
Event Type
Malfunction
Date Received
December 15, 2021
Date of Event
November 3, 2021
Report Date
December 1, 2021
Manufacturer
BARD ACCESS SYSTEMS, INC.
Product Code
LJS
Product Problem
Yes
Report Source
User Facility report
Reporter Location
UT
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Description of Event or Problem · 0

BARD ACCESS SYSTEMS, INC. 6.6F BROVIAC SINGLE LUMEN CATHETER WAS PLACES AND REPAIR TWO MONTHS AFTER. CHEST X-RAY (CXR) SHOWED REPAIR INTACT. FAMILY BROUGHT PATIENT INTO THE EMERGENCY DEPARTMENT (ED) TO EVALUATE POSITIONING AFTER LINE WAS CAUGHT ON CAR SEAT AS PATIENT GOT OUT AND IT LATER BECAME DIFFICULT TO FLUSH. CXR IN ED SHOWED THAT THE TIP REMAINED IN GOOD POSITION BUT REPAIR SITE SHOWED TRANSVERSE LATENCY WITHOUT STENT. REPAIR STENT APPEARS TO HAVE MIGRATED TO ~1.5CM FROM WHERE LINE ENTERS THE SKIN, REQUIRING LINE TO BE SURGICALLY REPLACED. THE STENT COMES FROM THE MANUFACTURER PERMANENTLY ATTACHED TO THE REPAIR SECTION OF THE REPAIR KIT, AND THE SLEEVE SLIDES OVER THE REPAIR SITE AND IS GLUED IN PLACE. WHILE THE FAMILY ORIGINALLY CAME IN WITH THE HISTORY OF THE LINE BEING CAUGHT IN A CAR SEAT, FOR THE STENT TO MIGRATE TO THE POSITION IT WAS IN, IT WOULD HAVE FIRST HAD TO BECOME DISLODGED FROM THE MANUFACTURER'S ATTACHMENT POINT WITHOUT THE REPAIR SITE SEPARATING AND THEN TRAVEL 14CM TOWARD THE PATIENT WITHIN 3.5 HOURS. IT'S ALSO HELD TOGETHER BY GLUE OUTSIDE THE LINE. BETWEEN THE LAYERS OF THE EXTERNAL CATHETER AND INTERNAL LUMEN OF THE REPAIR SLEEVE. THERE IS LITTLE, IF ANY GLUE THAT IS IN CONTACT WITH THE STENT AND THEREFORE A PULLING FORCE WOULD PUT SHEARING FORCE ON THE GLUE OF THE REPAIR BETWEEN THE LAYERS, AND NOT ON THE STENT ITSELF. UNKNOWN LOT NUMBER AT TIME OF REPAIR, CURRENT LOT NUMBER IN HOUSE IS REFQ0276.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1907030 BROVIAC CV CATHETERS (WHITE ADAPTER) CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS LJS BARD ACCESS SYSTEMS, INC. 0601620

Patients

Seq Age Sex Outcome Treatment
1 1825 DA Male Other