FDA Adverse Event Malfunction Summary report: N

TANGO M2

MDR report key: 13009674 · Received December 15, 2021

Report

Report Number
13009674
Event Type
Malfunction
Date Received
December 15, 2021
Date of Event
October 19, 2021
Report Date
December 1, 2021
Manufacturer
SUNTECH MEDICAL, INC.
Product Code
DXN
Product Problem
Yes
Report Source
User Facility report
Reporter Location
CA
Reporter Occupation
BIOMEDICAL ENGINEER
Health Professional
Yes

Narratives

Description of Event or Problem · 0

TANGO M2 STRESS TEST MONITOR FAILING TO PROVIDE PROPER BLOOD PRESSURE READING. "EXCESSIVE ARM MOVEMENT OR MICROPHONE NOISE." IS AN ERROR MESSAGE RECEIVED AFTER MULTIPLE TRIES, USING MULTIPLE CUFFS. THIS DEVICE (AUTOMATIC-INFLATION ELECTRONIC SPHYGMOMANOMETER) HAS FAILED IN THE PAST WITH THE SAME PROBLEM. VENDOR CHECKED IT AND REPORTED NO ISSUE FOUND. THE DEVICE WAS SENT AGAIN FOR THE SAME PROBLEM, AND THE UNIT WAS JUST RECEIVED AND STILL SHOWING THE SAME PROBLEM, EVEN THOUGH THE VENDOR SAYING NO PROBLEM WITH IT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1906791 TANGO M2 SYSTEM, MEASUREMENT, BLOOD-PRESSURE, NON-INVASIVE DXN SUNTECH MEDICAL, INC. 99-0088-51

Patients

Seq Age Sex Outcome Treatment
1 36135 DA Unknown