MDR report key: 13008736 · Received December 15, 2021

Report

Report Number
3003768277-2021-10274
Event Type
Malfunction
Date Received
December 15, 2021
Date of Event
November 19, 2021
Report Date
December 9, 2024
Manufacturer
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Product Code
OWB
UDI-DI
00884838085251
PMA / PMN Number
K172822
Removal / Correction Number
3003768277- 2021/10/29-0
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

THE WIRELESS CONNECTION WITH THE BASE STATION WAS RE-ESTABLISHED AFTER RESTARTING THE SYSTEM. PHILIPS HAS CONFIRMED THAT THE REPORTED FAILURE IS DUE TO A CONNECTIVITY ISSUE. DUE TO A FIRMWARE BUG THE WIRELESS FOOT SWITCH CAN SUDDENLY STOP RESPONDING WHEN A NUMBER OF AMBIENT CONDITIONS COEXIST, SUCH AS EMC DISTURBANCE AND THE PRESENCE OF OTHER WIRELESS DEVICES IN THE ROOM. PHILIPS HAS INITIATED A MEDICAL DEVICE CORRECTION/FIELD SAFETY CORRECTIVE ACTION (2021-IGT-BST-020). PHILIPS HAS ATTEMPTED TO OBTAIN PATIENT INFORMATION. THE CUSTOMER DID NOT RECALL THE PATIENT DETAILS. NO PATIENT INFORMATION COULD BE OBTAINED.

Additional Manufacturer Narrative · 0

THIS REPORT WAS SUBMITTED TO PROVIDE THE LINKAGE TO AN EXISTING CORRECTION & REMOVAL (3003768277- 2021/10/29-004-C). THE REPORTED EVENT OCCURRED PRIOR TO COMPLETION OF THE CORRECTION & REMOVAL 3003768277- 2021/10/29-004-C, WHICH ADDRESSES THE CAUSES ASSOCIATED WITH THE REPORTED MALFUNCTION.

Description of Event or Problem · 0

IT HAS BEEN REPORTED TO PHILIPS THAT DURING A PLANNED TREATMENT PROCEDURE THE WIRELESS FOOTSWITCH LOST CONNECTION WITH THE BASE STATION AND THE WIFI INDICATOR WAS RED. THE PROCEDURE WAS COMPLETED USING THE WIRED FOOTSWITCH. NO HARM TO THE PATIENT HAS BEEN REPORTED TO PHILIPS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1908300 AZURION INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM OWB PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. AZURION 7 M12 00884838085251

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown