COBAS 600 C501 MODULE
Report
- Report Number
- 1823260-2008-09424
- Event Type
- Malfunction
- Date Received
- December 23, 2008
- Date of Event
- December 6, 2008
- Report Date
- December 23, 2008
- Manufacturer
- ROCHE DIAGNOSTICS
- Product Code
- ---
- PMA / PMN Number
- K060373
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AR, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
USER EXPERIENCED THREE PT SAMPLES WITH DISCREPANT RESULTS FOR SODIUM AND OR BICARBONATE. INITIAL SODIUM GAVE 165 MMOL/L; REPEAT GAVE 134 MMOL/L. THE SODIUM RESULTS WERE NOT REPORTED, THEY WERE REPEATED PRIOR TO REPEATING RESULTS. THE FIELD SERVICE REPRESENTATIVE DETERMINED THE CAUSE TO BE FLUID DISPENSE WAS TOO HIGH ON CELL RINSE MECHANISM AND OUTSIDE SQUEE-GEE WAS GETTING HUNG UP INTERMITTENTLY. HE CLEANED SQUEE-GEE THEN ADJUSTED FLUID DISPENSE ON RINSE MECHANISM. PERFORMANCE TESTS WERE PERFORMED AND WITHIN SPECIFICATION.
USER EXPERIENCED THREE PT SAMPLES WITH DISCREPANT RESULTS FOR SODIUM AND OR BICARBONATE. IN 2008, INITIAL SODIUM GAVE 163 MMOL/L; REPEAT GAVE 134 MMOL/L. THE SODIUM RESULTS WERE NOT REPORTED, THEY WERE REPEATED PRIOR TO REPEATING RESULTS. THE FIELD SERVICE REPRESENTATIVE DETERMINED THE CAUSE TO BE FLUID DISPENSE WAS TOO HIGH ON CELL RINSE MECHANISM AND OUTSIDE SQUEE-GEE WAS GETTING HUNG UP INTERMITTENTLY. HE CLEANED SQUEE-GEE THEN ADJUSTED FLUID DISPENSE ON RINSE MECHANISM. PERFORMANCE TESTS WERE PERFORMED AND WITHIN SPECIFICATION.
USER EXPERIENCED THREE PT SAMPLES WITH DISCREPANT RESULTS FOR SODIUM AND OR BICARBONATE. INITIAL BICARBONATE GAVE 49 MMOL/L; REPEAT GAVE 23 MMOL/L. THE SODIUM RESULTS WERE NOT REPORTED, THEY WERE REPEATED PRIOR TO REPEATING RESULTS. THE FIELD SERVICE REPRESENTATIVE DETERMINED THE CAUSE TO BE FLUID DISPENSE WAS TOO HIGH ON CELL RINSE MECHANISM AND OUTSIDE SQUEE-GEE WAS GETTING HUNG UP INTERMITTENTLY. HE CLEANED SQUEE-GEE THEN ADJUSTED FLUID DISPENSE ON RINSE MECHANISM. PERFORMANCE TESTS WERE PERFORMED AND WITHIN SPECIFICATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COBAS 600 C501 MODULE | CLINICAL CHEMISTRY ANALYZER - CEM | --- | ROCHE DIAGNOSTICS | C501 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 64 YR | |||
| 2 | 70 YR | |||
| 3 | 66 YR |