FDA Adverse Event Malfunction Summary report: N

COBAS 600 C501 MODULE

MDR report key: 1300773 · Received December 23, 2008

Report

Report Number
1823260-2008-09424
Event Type
Malfunction
Date Received
December 23, 2008
Date of Event
December 6, 2008
Report Date
December 23, 2008
Manufacturer
ROCHE DIAGNOSTICS
Product Code
---
PMA / PMN Number
K060373
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AR, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Description of Event or Problem · 1

USER EXPERIENCED THREE PT SAMPLES WITH DISCREPANT RESULTS FOR SODIUM AND OR BICARBONATE. INITIAL SODIUM GAVE 165 MMOL/L; REPEAT GAVE 134 MMOL/L. THE SODIUM RESULTS WERE NOT REPORTED, THEY WERE REPEATED PRIOR TO REPEATING RESULTS. THE FIELD SERVICE REPRESENTATIVE DETERMINED THE CAUSE TO BE FLUID DISPENSE WAS TOO HIGH ON CELL RINSE MECHANISM AND OUTSIDE SQUEE-GEE WAS GETTING HUNG UP INTERMITTENTLY. HE CLEANED SQUEE-GEE THEN ADJUSTED FLUID DISPENSE ON RINSE MECHANISM. PERFORMANCE TESTS WERE PERFORMED AND WITHIN SPECIFICATION.

Description of Event or Problem · 2

USER EXPERIENCED THREE PT SAMPLES WITH DISCREPANT RESULTS FOR SODIUM AND OR BICARBONATE. IN 2008, INITIAL SODIUM GAVE 163 MMOL/L; REPEAT GAVE 134 MMOL/L. THE SODIUM RESULTS WERE NOT REPORTED, THEY WERE REPEATED PRIOR TO REPEATING RESULTS. THE FIELD SERVICE REPRESENTATIVE DETERMINED THE CAUSE TO BE FLUID DISPENSE WAS TOO HIGH ON CELL RINSE MECHANISM AND OUTSIDE SQUEE-GEE WAS GETTING HUNG UP INTERMITTENTLY. HE CLEANED SQUEE-GEE THEN ADJUSTED FLUID DISPENSE ON RINSE MECHANISM. PERFORMANCE TESTS WERE PERFORMED AND WITHIN SPECIFICATION.

Description of Event or Problem · 3

USER EXPERIENCED THREE PT SAMPLES WITH DISCREPANT RESULTS FOR SODIUM AND OR BICARBONATE. INITIAL BICARBONATE GAVE 49 MMOL/L; REPEAT GAVE 23 MMOL/L. THE SODIUM RESULTS WERE NOT REPORTED, THEY WERE REPEATED PRIOR TO REPEATING RESULTS. THE FIELD SERVICE REPRESENTATIVE DETERMINED THE CAUSE TO BE FLUID DISPENSE WAS TOO HIGH ON CELL RINSE MECHANISM AND OUTSIDE SQUEE-GEE WAS GETTING HUNG UP INTERMITTENTLY. HE CLEANED SQUEE-GEE THEN ADJUSTED FLUID DISPENSE ON RINSE MECHANISM. PERFORMANCE TESTS WERE PERFORMED AND WITHIN SPECIFICATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COBAS 600 C501 MODULE CLINICAL CHEMISTRY ANALYZER - CEM --- ROCHE DIAGNOSTICS C501

Patients

Seq Age Sex Outcome Treatment
1 64 YR
2 70 YR
3 66 YR