FDA Adverse Event Malfunction Summary report: N

TRUE METRIX

MDR report key: 13005753 · Received December 14, 2021

Report

Report Number
1000113657-2021-00775
Event Type
Malfunction
Date Received
December 14, 2021
Date of Event
November 17, 2021
Report Date
January 25, 2022
Manufacturer
TRIVIDIA HEALTH INC
Product Code
NBW
UDI-DI
00021292007836
PMA / PMN Number
K140100
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

SECTIONS WITH ADDITIONAL INFORMATION AS OF 25-JAN-2022: D9: DEVICE AVAILABLE FOR EVALUATION. H3: DEVICE EVALUATED BY MANUFACTURER. H6: UPDATED FDA¿S TYPE, FINDINGS AND CONCLUSIONS CODES. H10: TEST STRIPS WERE NOT RETURNED FOR EVALUATION. METER WAS RETURNED FOR EVALUATION. PRODUCT TESTING WAS PERFORMED AND NO DEFECT FOUND. MOST LIKELY UNDERLYING ROOT CAUSE: MLC-006: PASSED EXPIRATION DATE.

Additional Manufacturer Narrative · 0

INTERNAL REPORT REFERENCE NUMBER: (B)(4). METER AND TEST STRIPS WERE NOT RETURNED FOR EVALUATION. NOTE: MANUFACTURER CONTACTED CUSTOMER IN A FOLLOW-UP CALL ON 29-NOV-2021 TO ENSURE THE REPLACEMENT PRODUCTS RESOLVED THE INITIAL CONCERN - ABLE TO ESTABLISH CONTACT WITH CUSTOMER WHO STATED REPLACEMENT PRODUCTS RESOLVED INITIAL CONCERN.

Description of Event or Problem · 0

CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TEST RESULTS FROM RESULTS OBTAINED OF 203, 210, 229, 246 AND 267MG/DL. THE CUSTOMER¿S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 75-130MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL, A BACK TO BACK BLOOD TEST WAS NOT PERFORMED BY THE CUSTOMER. THE PRODUCT IS STORED ACCORDING TO SPECIFICATION IN THE BEDROOM. THE TEST STRIP LOT MANUFACTURER¿S EXPIRATION DATE IS 01/31/2023 AND TEST STRIPS WERE OPENED (B)(6) 2021 (EXPIRED). THE CUSTOMER DID NOT HAVE ANOTHER VIAL OF TEST STRIPS THAT HAD BEEN STORED AND HANDLED CORRECTLY. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: (B)(6).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1905084 TRUE METRIX SYSTEM, TEST BLOOD GLUCOSE, OVER THE COUNTER NBW TRIVIDIA HEALTH INC STRIP, LEADER TMX 50CTMM MG/DL MY4441S 00021292007836

Patients

Seq Age Sex Outcome Treatment
1 Unknown