FDA Adverse Event Injury Summary report: N

TACTICATH¿ CONTACT FORCE ABLATION CATHETER, SENSOR ENABLED¿ BID CURVE D-F

MDR report key: 13001703 · Received December 14, 2021

Report

Report Number
3008452825-2021-00557
Event Type
Injury
Date Received
December 14, 2021
Date of Event
November 25, 2021
Report Date
December 14, 2021
Manufacturer
ST. JUDE MEDICAL
Product Code
OAE
UDI-DI
05415067027641
PMA / PMN Number
P130026
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE RESULTS OF THE INVESTIGATION ARE INCONCLUSIVE SINCE THE DEVICE WAS NOT RETURNED FOR ANALYSIS; HOWEVER, THE LOG FILES FROM THE TACTISYS QUARTZ SYSTEM WERE RETURNED. THE LOG FILE ANALYSIS CONCLUDED THAT THE TACTICATH CATHETER PERFORMED AS INTENDED. THE OPTICAL FIBERS MET SPECIFICATIONS, THE RECORDED TEMPERATURES INDICATED COOLING DURING RF ABLATION, AND CONTACT FORCE MEASUREMENTS WERE DISPLAYED THROUGHOUT THE DURATION OF THE LOG FILES. THE DEVICE HISTORY RECORD WAS REVIEWED TO ENSURE THAT EACH MANUFACTURING AND INSPECTION OPERATION WAS PERFORMED. THE CAUSE OF THE REPORTED EVENT REMAINS UNKNOWN.

Description of Event or Problem · 0

DURING AN ATRIAL FIBRILLATION PROCEDURE, A CARDIAC TAMPONADE WAS OBSERVED. WHILE THE ABLATION CATHETER WAS IN THE LEFT ATRIUM (LA), AT THE ANTERIOR WALL LOOKING FOR THE CFE, THE CATHETER APPEARED SLIGHTLY OUTSIDE THE GEOMETRY WITH AN IMPEDANCE SLIGHTLY HIGHER THAN THE LA. THE CATHETER WAS MOVED, BUT NOT POSTERIOR. THE CATHETER WAS THEN MOVED A FEW CENTIMETERS AND THE CATHETER WAS BACK IN THE LA WITH GOOD IMPEDANCE. A FEW MINUTES LATER, THE PATIENT BECAME HYPOTENSIVE, AND A CARDIAC TAMPONADE WAS REVEALED VIA TRANSOESOPHAGEAL ECHO (ETO) AND TRANS THORACIC ECHO (ETT). THE BLOOD WAS DRAINED AND THE PATIENT WAS STABILIZED. THERE WERE NO PERFORMANCE ISSUES WITH ANY ABBOTT DEVICES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1898578 TACTICATH¿ CONTACT FORCE ABLATION CATHETER, SENSOR ENABLED¿ BID CURVE D-F CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION OAE ST. JUDE MEDICAL A-TCSE-DF 8001682 05415067027641

Patients

Seq Age Sex Outcome Treatment
1 59 YR Female Required Intervention