FDA Adverse Event Malfunction Summary report: N

BD INTIMA II¿ IV CATHETER WITH PRN ADAPTER

MDR report key: 13001688 · Received December 14, 2021

Report

Report Number
3014704491-2021-00351
Event Type
Malfunction
Date Received
December 14, 2021
Date of Event
November 3, 2021
Report Date
November 29, 2021
Manufacturer
BECTON DICKINSON MEDICAL DEVICES CO., LTD. SUZHOU
Product Code
FOZ
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

(B)(6). INVESTIGATION SUMMARY: IN RESPONSE TO THE EVENT REPORTED BY YOUR FACILITY A DEVICE HISTORY REVIEW WAS CONDUCTED FOR LOT NUMBER 1020006. OUR RECORDS SHOW THAT THIS IS THE ONLY INSTANCE OF THIS ISSUE OCCURRING IN THIS PRODUCTION BATCH. ACCORDING TO THE SAMPLING PLAN APPLIED FOR PRODUCT PERFORMANCE, THIS LOT WAS ACCEPTED AND RELEASED WITHOUT DEFECTS BEING NOTED DURING THE FINAL ASSEMBLY OR VISUAL INSPECTIONS. A SAMPLE COULD NOT BE OBTAINED FOR EVALUATION AND TESTING; IN LIEU OF THE AFFECTED DEVICE, FUNCTIONAL TESTING WAS PERFORMED ON RETENTION SAMPLES FOR THIS LOT, THE RESULTS OF THESE SHOW THAT THE TESTED UNITS PERFORMED WITHIN PRODUCT SPECIFICATIONS. UNFORTUNATELY WITHOUT THE ABILITY TO INVESTIGATE THE AFFECTED UNIT OUR QUALITY ENGINEERS WERE UNABLE TO DETERMINE THE ROOT CAUSE FOR THIS COMPLAINT.

Description of Event or Problem · 0

IT WAS REPORTED WHEN USING THE BD INTIMA II¿ IV CATHETER WITH PRN ADAPTER, THE DEVICE EXPERIENCED LEAKAGE AT THE CATHETER AND HUB JUNCTION. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER. THE CUSTOMER STATED: AT 9:10 AM ON (B)(6) 2021, WHEN THE NURSE WAS CONDUCTING INDWELLING NEEDLE VENIPACTURE FOR THE PATIENT'S INFUSION, BLOOD WAS RETURNED AFTER THE INFUSION, AND BLOOD LEAKAGE WAS FOUND AT THE CATHETER TUBE CONNECTOR BEHIND THE INDWELLING NEEDLE, SO SHE IMMEDIATELY REMOVED AND REPLACED THE INDWELLING NEEDLE FOR RE-PUNCTURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1898172 BD INTIMA II¿ IV CATHETER WITH PRN ADAPTER INTRAVASCULAR CATHETER FOZ BECTON DICKINSON MEDICAL DEVICES CO., LTD. SUZHOU 1020006

Patients

Seq Age Sex Outcome Treatment
1 Female