FDA Adverse Event Malfunction Summary report: N

SINGLE USE 3-LUMEN SPHINCTEROTOME V

MDR report key: 12999917 · Received December 14, 2021

Report

Report Number
8010047-2021-16081
Event Type
Malfunction
Date Received
December 14, 2021
Date of Event
November 19, 2021
Report Date
December 14, 2021
Manufacturer
OLYMPUS MEDICAL SYSTEMS CORP.
Product Code
KNS
PMA / PMN Number
K950166
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
NURSE
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE SUBJECT DEVICE WAS EVALUATED AND INVESTIGATED BY OLYMPUS MEDICAL SYSTEMS CORP. (OMSC). THE CUTTING WIRE WAS BROKEN. INVESTIGATION WAS CARRIED OUT TO CONFIRM THE BROKEN PORTION. THE COATED PORTION OF THE CUTTING WIRE WAS TORN, AND THE BROKEN PORTION WAS SCORCHED AND MELTED. THE OUTER DIAMETER OF THE CUTTING WIRE WAS MEASURED. THE RESULT INDICATED NO ABNORMALITIES. THE LENGTH OF THE COATED PORTION OF THE CUTTING WIRE, AND THE CUTTING WIRE ITSELF PRESENTED NO ABNORMALITIES. THERE WERE NO MISSING PARTS IN THE SUBJECT DEVICE. OTHER ABNORMALITIES THAT COULD LEAD TO THE BREAKAGE OF THE CUTTING WIRE WERE NOT CONFIRMED. THE FOLLOWING INFORMATION WAS PROVIDED FROM THE INITIATOR. ¿WHAT TYPE OF THE PROCEDURE WAS BEING PERFORMED? (WHICH PART OF THE BODY?) ;UNKNOWN ¿WHICH ELECTROSURGICAL UNIT WAS BEING USED FOR THE PROCEDURE? ;ESG-100 ¿WHAT WAS THE SETTING VALUE FOR THE CUT MODE AND THE COAGULATION MODE? ;UNKNOWN ¿AT WHAT TIMES DID THE REPORTED EVENT OCCUR? (FOR EXAMPLE: HOW MANY MINUTES AFTER THE PROCEDURE STARTED, DID THE REPORTED EVENT OCCUR?) ;UNKNOWN ¿WHICH MODE WAS BEING USED WHEN THE EVENT OCCURRED? (CUT, BLEND OR COAGULATION) ;UNKNOWN ¿HOW WAS THE WIRE CONTACTING THE TISSUE WHEN THE REPORTED EVENT OCCURRED? ;UNKNOWN ¿WHAT IS THE AVERAGE ACTIVATION TIME FOR THE CUT MODE OR THE COAGULATION MODE? HOW LONG IS THE MAXIMUM ACTIVATION TIME? ;UNKNOWN. ¿OTHER INFORMATION TO HELP PERFORMING AN INVESTIGATION WHILE HANDLING THE DEVICE (DURING THE PREPARATION OR THE PROCEDURE);- NO ABNORMALITIES WERE DETECTED IN THE DEVICE HISTORY RECORD WITH THE LOT NUMBER FOR THE FOLLOWING INSPECTION ITEMS WHICH RELATED TO THE REPORTED PHENOMENON. ¿LENGTH OF CUTTING WIRE ¿LENGTH OF COATED PORTION ¿OPERATION OF CUTTING WIRE BASED ON THE CONFIRMATION RESULT AND THE INVESTIGATION RESULTS IN THE PAST, A LIKELY MECHANISM CAUSING THE BROKEN CUTTING WIRE MIGHT BE THE FOLLOWING. THE CUTTING WIRE AT A TORN AREA OF THE COATED PORTION CAME INTO CONTACT WITH THE DISTAL END OF THE ENDOSCOPE WHILE THE FORCEPS ELEVATOR WAS RAISED. THE OUTPUT WAS ACTIVATED IN STATE OF ¿1¿ DESCRIPTION, AND THE CUTTING WIRE BECAME HOT INSTANTLY. THIS CAUSED THE CUTTING WIRE TO BREAK. IT HAS BEEN CONFIRMED THAT THE TEAR OF THE COATED PORTION OF THE CUTTING WIRE COULD DUPLICATE BY THE FOLLOWING MECHANISM. THE FORCEPS ELEVATOR OF THE ENDOSCOPE WAS RAISED. WHEN THE CUTTING WIRE DEFLECTS, THE COATED PORTION OF THE CUTTING WIRE AND THE METAL PART OF THE DISTAL END OF THE ENDOSCOPE COME INTO CONTACT. IN STATE OF DESCRIPTION ¿2¿, THE CUTTING WIRE WAS MOVED BACK AND FORTH. THIS CAUSED THE COATED PORTION OF THE CUTTING WIRE TO TEAR. THE SLIDER WAS PUSHED MORE THAN NEEDED. THIS CAUSED THE CUTTING WIRE TO DEFLECT. THE ABOVE DEVICE HANDLING HAS WARNED IN THE INSTRUCTION MANUAL.

Description of Event or Problem · 0

OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) WAS INFORMED BY A NURSE THAT DURING AN ENDOSCOPIC RETROGRADE COLANGIOPANCREATOGRAPHY USING THE SUBJECT DEVICE WITH ESG-100, THE CUTTING WIRE WAS BROKEN ON ACTIVATION. THEN, THE IMAGE OF THE ENDOSCOPE LOOKED NOISY, AND A SURGEON COULD NOT SHOW. THE INTENDED PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. THERE WAS NO PATIENT INJURY REPORTED. THE SUBJECT DEVICE WAS RETURNED TO OMSC FOR EVALUATION AND INVESTIGATION. OMSC FOUND THAT NOT ONLY THE CUTTING WIRE WAS BROKEN BUT ALSO THE COATED PORTION OF THE CUTTING WIRE WAS TORN, AND THE BROKEN PORTION WAS SCORCHED AND MELTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1900043 SINGLE USE 3-LUMEN SPHINCTEROTOME V SINGLE USE 2-LUMEN SPHINCTEROTOME KNS OLYMPUS MEDICAL SYSTEMS CORP. KD-V411M-0725 16K

Patients

Seq Age Sex Outcome Treatment
1 Unknown