FDA Adverse Event
Malfunction
Summary report: N
NUCLETRON PLATO BPS
MDR report key: 1299165
·
Received December 4, 2008
Report
- Report Number
- 9611894-2008-00015
- Event Type
- Malfunction
- Date Received
- December 4, 2008
- Date of Event
- July 18, 2008
- Report Date
- October 14, 2008
- Manufacturer
- NUCLETRON BV
- Product Code
- MUJ
- PMA / PMN Number
- K9833343
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
IN JULY 2008, A BULLETIN DESCRIBING THE CONDITION WHEN INCORRECT DOSE VALUES MIGHT BE DISPLAYED WAS DISTRIBUTED TO USERS OF PLATO BPS V14.3 - 14.7.
Description of Event or Problem · 1
WHEN CREATING A PLAN IN PLATO BPS VERSION 14.3.3, 14.3.5, 14.3.6, AND 14.3.7 INCORRECT DOSE PERCENTAGES ARE BEING DISPLAYED IN THE POINTS WINDOW WHEN CREATING OF MOVING POINTS IN A CT SLICE. DOSE IN PATIENT POINTS SHOULD BE CORRECTLY CALCULATED AND DISPLAYED IF A PATIENT POINT IS ADDED OR MOVED IN A CT IMAGE. AS A RESULT, PLANS CAN BE INCORRECTLY EVALUATED UNLESS RE-PRESCRIPTION IS PERFORMED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | NUCLETRON PLATO BPS | TREATMENT PLANNING | MUJ | NUCLETRON BV | 14.3.3 - 14.3.7 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |