FDA Adverse Event Malfunction Summary report: N

NUCLETRON PLATO BPS

MDR report key: 1299165 · Received December 4, 2008

Report

Report Number
9611894-2008-00015
Event Type
Malfunction
Date Received
December 4, 2008
Date of Event
July 18, 2008
Report Date
October 14, 2008
Manufacturer
NUCLETRON BV
Product Code
MUJ
PMA / PMN Number
K9833343
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

IN JULY 2008, A BULLETIN DESCRIBING THE CONDITION WHEN INCORRECT DOSE VALUES MIGHT BE DISPLAYED WAS DISTRIBUTED TO USERS OF PLATO BPS V14.3 - 14.7.

Description of Event or Problem · 1

WHEN CREATING A PLAN IN PLATO BPS VERSION 14.3.3, 14.3.5, 14.3.6, AND 14.3.7 INCORRECT DOSE PERCENTAGES ARE BEING DISPLAYED IN THE POINTS WINDOW WHEN CREATING OF MOVING POINTS IN A CT SLICE. DOSE IN PATIENT POINTS SHOULD BE CORRECTLY CALCULATED AND DISPLAYED IF A PATIENT POINT IS ADDED OR MOVED IN A CT IMAGE. AS A RESULT, PLANS CAN BE INCORRECTLY EVALUATED UNLESS RE-PRESCRIPTION IS PERFORMED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NUCLETRON PLATO BPS TREATMENT PLANNING MUJ NUCLETRON BV 14.3.3 - 14.3.7

Patients

Seq Age Sex Outcome Treatment
1