MICROSELECTRON HDR V2
Report
- Report Number
- 9611894-2008-00016
- Event Type
- Malfunction
- Date Received
- December 4, 2008
- Date of Event
- February 13, 2008
- Report Date
- October 14, 2008
- Manufacturer
- NUCLETRON BV
- Product Code
- JAQ
- PMA / PMN Number
- K953946
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE PROBLEM INVESTIGATION INCLUDED A REVIEW OF THE FACILITY'S TREATMENT LOGS, THE TREATMENT CONTROL PANEL (TCP) FIRMWARE CODE, AND TREATMENT CONTROL STATION (TCS) SOFTWARE. IT COULD NOT BE DETERMINED WHAT CAUSED THE COMPUTER CRASH RESULTING IN A COMMUNICATION ERROR BETWEEN THE TREATMENT UNIT (TU) AND THE TREATMENT CONTROL PANEL (TCP) DURING TREATMENT. MOST LIKELY, THERE WAS A HARDWARE FAILURE IN THE PC WHICH CAUSED THE LOSS OF COMMUNICATION. IN THIS SITUATION, THE TCP PRINTOUT SHOULD PROVIDE A DETAILED TREATMENT RECORD. HOWEVER, AT THIS SITE THE TCP PRINTER PORT WAS NOT CONNECTED TO THE PRINTER. WHEN SERVICE REPLACED THE PC AND RECONNECTED THE TCP, THE PRINTOUT STILL COULD NOT BE MADE. INVESTIGATION OF THE SOFTWARE CODE SHOWS THAT THE TCP DATA IS ONLY RESET WHEN NEW TREATMENT DATA IS SENT FROM THE TCS OR WHEN A CHECK/SOURCE CABLE EXCHANGE AND CALIBRATION MODE IS SELECTED. MOST LIKELY THE TCP DATA WAS FLUSHED WHEN THE FACILITY TRIED TO REBOOT THE SYSTEM AND A PARTIAL RECOVERY WAS DONE. FULL RECOVERY WAS NOT APPARENTLY POSSIBLE BECAUSE OF THE PROBLEM WITH THE PC. THE TCP PRINTER CONNECTION SHOULD HAVE BEEN INSTALLED AND OPERATIONAL. INDEPENDENT SECONDARY TIMES KEPT BY USERS DURING TREATMENT PROVIDE A SAFEGUARD AT THE USER INTERFACE. NUCLETRON ISSUED A BULLETIN TO ALL CUSTOMERS THAT ADDRESSED THE IMPORTANCE OF THE EMERGENCY PRINTOUT AND ALTERNATIVE WAYS TO RETRIEVE TREATMENT DATA FROM THE AFTERLOADER IF THE PRINTER FAILS.
CPU CRASHED IN TREATMENT. WHEN PC REBOOTED THERE WAS NO TREATMENT DATA FOR THE TREATMENT. CUSTOMER CALLED SUPPORT LINE. TECH SUPPORT TRIED TO HAVE CUSTOMER PRESS THE PRINT BUTTON ON THE TCP. CUSTOMER PRESSED PRINT, BUT NOTHING HAPPENED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MICROSELECTRON HDR V2 | HDR REMOTE AFTERLOADER | JAQ | NUCLETRON BV |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |