FDA Adverse Event Malfunction Summary report: N

ONCENTRA MASTERPLAN

MDR report key: 1299161 · Received December 4, 2008

Report

Report Number
9611894-2008-00019
Event Type
Malfunction
Date Received
December 4, 2008
Date of Event
September 2, 2008
Report Date
October 14, 2008
Manufacturer
NUCLETRON BV
Product Code
MUJ
PMA / PMN Number
K081281
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Description of Event or Problem · 1

OMP BP ADDS A FRACTION AUTOMATICALLY WHILE TRANSFERRING THE PLAN TO ONCENTRA TCS. ONE OF OUR CUSTOMERS USING ONCENTRA BRACHY VERSION 3.1SP1 WITH ONCENTRA TCS VERSION 3.1.2.100 HAS REPORTED A PROBLEM. HE PERFORMED A PLAN AND EXPORTED IT TO THE TCS WITH ONLY (1) ONE FRACTION. FOR SOME REASON HE WANTED TO CHANGE THE NUMBER OF FRACTIONS SO HE COPIED THE PLAN AND CHANGED ONLY THE NUMBER OF FRACTIONS TO 3 AND RE-EXPORTED THE SAME PLAN WITH A DIFFERENT NAME TO TCS. ON IMPORTING THE SECOND STUDY IN THE TCS IT NOW SHOWS THE NUMBER OF FRACTIONS AS (4) FOUR INSTEAD OF 3. THIS PROBLEM COMES ONLY WHEN YOU COPY THE PLAN ON ONCENTRA MASTERPLAN, IF YOU DIRECTLY CHANGE THE NUMBER OF FRACTIONS IN THE SAME PLAN INSTEAD OF COPYING THE PLAN IT IS SHOWING CORRECTLY.