FDA Adverse Event Malfunction Summary report: N

PENTAX

MDR report key: 12982044 · Received December 13, 2021

Report

Report Number
9610877-2021-50803
Event Type
Malfunction
Date Received
December 13, 2021
Date of Event
December 6, 2021
Report Date
December 13, 2021
Manufacturer
HOYA CORPORATION PENTAX TOKYO OFFICE
Product Code
FDF
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

MODEL EC38-I10L-US IS AVAILABLE IN THE USA WITH A 510K NUMBER K131855. WE CHECKED THE RETURNED UNIT AND CONFIRMED THE CCD MODULE AS DEFECTIVE. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE EXCESSIVE FORCE APPLIED ON THE CCD MODULE. BASED ON THE TECHNICAL REPORT AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.

Description of Event or Problem · 0

THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1881519 PENTAX VIDEO COLONOSCOPE FDF HOYA CORPORATION PENTAX TOKYO OFFICE EC38-I10NL

Patients

Seq Age Sex Outcome Treatment
1 Unknown