FDA Adverse Event
Malfunction
Summary report: N
PENTAX
MDR report key: 12982043
·
Received December 13, 2021
Report
- Report Number
- 9610877-2021-50790
- Event Type
- Malfunction
- Date Received
- December 13, 2021
- Date of Event
- December 7, 2021
- Report Date
- December 13, 2021
- Manufacturer
- HOYA CORPORATION PENTAX TOKYO OFFICE
- Product Code
- FDF
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
Additional Manufacturer Narrative · 0
MODEL EC38-I10L-US IS AVAILABLE IN THE USA WITH A 510K NUMBER K131855. WE CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE CCD MODULE DEFECTIVE. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE EXCESSIVE FORCE APPLIED ON THE CCD MODULE. BASED ON THE TECHNICAL REPORT AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.
Description of Event or Problem · 0
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1881522 | PENTAX | VIDEO COLONOCOPE - I10 STANDARD | FDF | HOYA CORPORATION PENTAX TOKYO OFFICE | EC38-I10L |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |