FDA Adverse Event Malfunction Summary report: N

ID NOW COVID-19 ASSAY

MDR report key: 12981410 · Received December 10, 2021

Report

Report Number
1221359-2021-03685
Event Type
Malfunction
Date Received
December 10, 2021
Report Date
January 19, 2022
Manufacturer
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
Product Code
QJR
UDI-DI
10811877011320
PMA / PMN Number
EUA2000074
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ID
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT M162544 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 191-000 / LOT M162544 AND TEST BASE PART NUMBER 190-430 / LOT M162544.THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT M162544 SHOWED THAT THE COMPLAINT RATE IS (B)(4). ABBOTT DIAGNOSTICS SCARBOROUGH WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE AS THE LOGFILES WERE NOT PROVIDED. HOWEVER, A POSSIBLE ASSIGNABLE ROOT CAUSE IS PATIENT SAMPLE INTERFERENCE. SUBSTANCES IN THE SAMPLE ITSELF MAY HAVE INTERFERED WITH THE REACTION.

Additional Manufacturer Narrative · 0

THE INVESTIGATION IS STILL IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION. PLEASE SEE MFR. REPORT: 1221359-2021-03683, 1221359-2021-03684 AND 1221359-2021-03685.

Description of Event or Problem · 0

THE CUSTOMER REPORTED (3) THREE FALSE POSITIVE RESULTS WITH THE ID NOW COVID-19 ASSAY PERFORMED ON UNKNOWN SWAB SAMPLES WITH TWO LOT NUMBERS (LOT. M160774. AND LOT M162544) . THIS MFR. REPORT ADDRESSES RESULTS 2 OF 3. THE CUSTOMER REPORTED THAT THE PATIENT TOOK THE ID NOW COVID-19 ASSAY AND CONTAINED A POSITIVE RESULT. PCR CONFIRMATION TESTING WAS PERFORMED AND A NEGATIVE RESULT WAS OBTAINED. NO ADDITIONAL PATIENT INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1877488 ID NOW COVID-19 ASSAY MOLECULAR IVD FOR ID NOW COVID-19 QJR ABBOTT DIAGNOSTICS SCARBOROUGH, INC. M162544 10811877011320

Patients

Seq Age Sex Outcome Treatment
1 Unknown