FDA Adverse Event Malfunction Summary report: N

FLEXLAB

MDR report key: 12974604 · Received December 10, 2021

Report

Report Number
3010825766-2021-00011
Event Type
Malfunction
Date Received
December 10, 2021
Report Date
February 24, 2022
Manufacturer
INPECO SA
Product Code
CEM
UDI-DI
07640172340004
PMA / PMN Number
K121012
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THE INVESTIGATION IS ONGOING TO DETERMINE THE ROOT CAUSE AND TO IDENTIFY HOW TO FIX THE ISSUE. THE WBB MODULES ARE KEPT MONITORED.

Additional Manufacturer Narrative · 0

THE INVESTIGATION IS ONGOING TO DETERMINE THE ROOT CAUSE AND TO IDENTIFY HOW TO FIX THE ISSUE. THE WBB MODULES ARE KEPT MONITORED.

Additional Manufacturer Narrative · 0

THE INITIAL REPORT HAS BEEN SUBMITTED ON DECEMBER 10TH, 2021. ADDITIONAL INFORMATION: THE INVESTIGATION HAS BEEN COMPLETED. THE ROOT CAUSE OF THE SAMPLE SPILLAGE AT THE WIDE BELT BUFFER (WBB) EXIT HAS BEEN IDENTIFIED IN THE SPECIFIC TRAFFIC CONDITIONS, WORKFLOW AND LAYOUT CONFIGURATION PRESENT IN THIS SITE. TO SOLVE THE PROBLEM THE SPEED VALUES OF THE WBB BELTS WHICH CARRY THE CARRIERS FROM THE BUFFER TO THE MODULE PIT LANE HAVE BEEN MODIFIED TO IMPROVE THE CONFLUENCE OF THE CARRIERS ON THE TRACK. AFTER THIS CHANGE THE MODULES HAVE BEEN KEPT MONITORED AND NO FURTHER OCCURRENCES HAVE BEEN OBSERVED WHERE THE SAMPLE TUBES RETURN ON THE MODULE PIT LANE. NO ADDITIONAL ACTIONS ARE FORESEEN.

Description of Event or Problem · 0

SAMPLE RESIDUES HAVE BEEN FOUND AT THE EXIT OF THE WIDE BELT CARRIER BUFFER MODULES (WBB) WHERE THE SAMPLE TUBES PREVIOUSLY PARKED IN THE MODULE BUFFER RETURNED ON THE PIT LANE. THE WIDE BELT CARRIER BUFFER MODULE IS A SAMPLE CARRIERS COLLECTOR THAT PROVIDES THE CAPABILITY TO RELIEVE THE TRACK WORKLOAD. THE SAMPLE SPILLAGE MAY HAVE CAUSED THE CONTAMINATION OF OTHER UNCAPPED SAMPLE TUBES PRESENT IN THE AREA. SO FAR NO CASES OF ACTUAL CROSS CONTAMINATION HAVE BEEN IDENTIFIED.

Description of Event or Problem · 0

SAMPLE RESIDUES HAVE BEEN FOUND AT THE EXIT OF THE WIDE BELT CARRIER BUFFER MODULES (WBB) WHERE THE SAMPLE TUBES PREVIOUSLY PARKED IN THE MODULE BUFFER RETURNED ON THE PIT LANE. THE WIDE BELT CARRIER BUFFER MODULE IS A SAMPLE CARRIERS COLLECTOR THAT PROVIDES THE CAPABILITY TO RELIEVE THE TRACK WORKLOAD. THE SAMPLE SPILLAGE MAY HAVE CAUSED THE CONTAMINATION OF OTHER UNCAPPED SAMPLE TUBES PRESENT IN THE AREA. SO FAR NO CASES OF ACTUAL CROSS CONTAMINATION HAVE BEEN IDENTIFIED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1874426 FLEXLAB LABORATORY AUTOMATION SYSTEM CEM INPECO SA FLX 07640172340004

Patients

Seq Age Sex Outcome Treatment
1 Unknown