670G INSULIN PUMP MMT-1780KL
Report
- Report Number
- 2032227-2021-228886
- Event Type
- Malfunction
- Date Received
- December 10, 2021
- Date of Event
- July 29, 2021
- Report Date
- December 10, 2021
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OZP
- UDI-DI
- 000000763000365882
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE PUMP WAS RETURNED FOR UNEXPECTED BATTERY POWER LOSS OR REPLACE BATTERY ALERT/REPLACE BATTERY NOW ALARM FOUND ON (B)(6) 2021. THE PUMP PASSED THE SELF TEST, DISPLACEMENT TEST AND ACTIVE CURRENT MEASUREMENT. HOWEVER, THE PUMP HAD A HIGH SLEEP CURRENT MEASUREMENT. SUCCESSFULLY DOWNLOADED HISTORY FILES AND TRACES USING THUS. POWER MANAGEMENT GRAPH WAS SUCCESSFULLY GENERATED. THE LOADED VOLTAGE (LOADED VLITH) AND THE UNLOADED VOLTAGE (UNLOADED VLITH) DISPLAY AT THE POWER GRAPH MANAGEMENT TOOL SHOWS ABNORMAL BEHAVIOR. NO ALARMS OR ALERTS NOTED DURING THE TESTING. HOWEVER, LOW BATTERY ALERT WAS RECORDED AND FOUND IN THE FORMATTED HISTORY FILE ON: (B)(6) 2021 15:48:00.000 ALARMALERTNOTIFICATION. 07/29/2021 16:03:31.000 ALARMALERTCLEARED. AND POWER LOSS ALARM WAS RECORDED AND FOUND IN THE FORMATTED HISTORY FILE ON: (B)(6) 2021 16:11:46.000 ALARMALERTNOTIFICATION. (B)(6) 2021 16:11:57.000 ALARMALERTNOTIFICATION. (B)(6) 2021 16:12:17.000 ALARMALERTCLEARED. PUMP WAS CUT OPEN TO PERFORM VISUAL INSPECTION AND FOUND NO EVIDENCE OF PHYSICAL OR MOISTURE DAMAGE ON THE PCBA1, PCBA2, FORCE SENSOR, MOTOR AND VIBRATOR ASSEMBLY NOTED. TEST P-CAP AND RESERVOIR LOCKED PROPERLY INTO RESERVOIR COMPARTMENT DURING TESTING. THE FOLLOWING WERE NOTED DURING VISUAL INSPECTION: A CRACKED BATTERY TUBE THREADS, A SCRATCHED CASE, A CRACKED CASE-CORNER OF BELT CLIP RAILS NEAR THE BATTERY TUBE COMPARTMENT AND A SERIAL NUMBER LABEL FADING. THE ORIGINAL PCB1 WAS INSTALLED IN A TEST PCB2, CASE, INTERNAL BATTERY AND MOTOR. POWERED THE PUMP ON USING THE BATTERY SIMULATOR AND CONTINUED TESTING. THE PUMP FAILED THE SLEEP CURRENT MEASUREMENT. THE ORIGINAL PCB2 WAS INSTALLED IN A TEST PCB1, CASE, INTERNAL BATTERY AND MOTOR. POWERED THE PUMP ON USING THE BATTERY SIMULATOR AND CONTINUED TESTING. THE PUMP PASSED THE SLEEP CURRENT MEASUREMENT. THE PUMP HAD A HIGH SLEEP CURRENT MEASUREMENT, PROBLEM ISOLATED ON THE PCB1. UNEXPECTED BATTERY POWER LOSS OR REPLACE BATTERY ALERT/REPLACE BATTERY NOW ALARM CONFIRMED. UNEXPECTED BATTERY POWER LOSS OR REPLACE BATTERY ALERT/REPLACE BATTERY NOW ALARM, PROBLEM ISOLATED ON THE PCB1.
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP RECEIVED REPLACE BATTERY ALARM. THE CUSTOMER RECEIVE A LOW BATTERY ALERT PRIOR TO THE REPLACE BATTERY ALERT. CUSTOMER STATED THE BATTERY CAP WAS NOT LOOSE, CRACKED OR DAMAGED. THIS WAS THE SECOND OCCURRENCE OF REPLACE BATTERY ALERT OR REPLACE BATTERY NOW WITHOUT A LOW BATTERY WARNING. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1879404 | 670G INSULIN PUMP MMT-1780KL | ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL | OZP | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-1780KL | HG501TZ | 000000763000365882 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 40 YR | Female |